6.1 Spun Cylinders. Spun cylinders may be installed in a
horizontal position on the DynaPak optional BackRack
vessel rack . Avoiding traps in the line, install stainless
steel tubing and fittings from the sample discharge port of
the sampler to the product end of the sample cylinder.
300cc and 500cc spun cylinders may also be installed in a
vertical position. Piping from the sampler discharge port to
the sample vessel should be arranged such that liquid
traps are not created.
6.2 Variable volume/constant pressure cylinder.
A free-floating piston cylinder (DuraSite) may be installed in
a horizontal position on an optional vessel rack. Free-
floating piston cylinders should NOT be installed on the
DynaPak BackRack vessel rack.
Install 1/8" tubing from the sample discharge port of the
manifold to the product end of the vessel. Avoid traps in this
line.
See diagram #8 for DuraSite portable sample vessel
instructions.
6.3 LinkPlus. Install the LinkPlus directly into the sample
discharge port of the sampler. Use stainless steel tubing
and fittings to connect the LinkPlus outlet to the product end
of the sample cylinder.
6. Sample Vessel Installation