background image

34

Diagram #9 - DuraSite Portable Sample Vessel Instructions

Purpose:  The DuraSite Portable Sample Vessel permits
the user to remove a liquid or gas
hydrocarbon sample from a pipeline or a sampling device.
This is accomplished without changing the pressure of the
product or exposing it to a
contaminant fluid.  If properly used and maintained the
DuraSite will provide many years of safe,
accurate and clean sampling.

Use:  The DuraSite is a very safe device to use.  As with any
equipment dealing with flammable
products, it is mandatory that a good, thorough operator
training procedure be established prior to use.

Typical use of the cylinder would be as follows:

Step 1:  (In The Lab)  Connect a regulated inert gas supply
to the pre-charge valve.  The product valve should be open.
By carefully controlling the
pre-charge valve and the regulator, the cylinder can be
slowly charged with pre-charge gas (NOTE:  This should be
done slowly to prevent slamming the piston down to the
opposite end).  The pressure on the pre-charge pressure
gauge should be brought to a reading of 10-50 psi above the
expected pressure of the product in the field .  Close the pre-
charge valve and disconnect the gas supply.  Check the
pre-charge valve, relief device, and the pre-charge pressure
gauge for leaks.  Any leaks should be stopped before
continuing.  The vessel should be placed in a padded
carrying case and made ready for field use.
Proceed to 

EITHER Step 2, or Step 3 as

required for your application.

STEP 2: FOR COLLECTION OF SAMPLE VIA SPOT
SAMPLE  OR  FROM  COMPOSITE
ACCUMULATOR  VESSEL.

2a:  Connect the product end of the pre-charged sample
vessel to the product supply.
(Sampler product removal valve, or Pipeline sample probe)
NOTE:  the pre-charge pressure gauge reading should
be greater than the product supply
pressure reading.  If not, repeat Step 1 above.

2b:  Once the vessel is connected to the product supply, it
is necessary to vent a small amount of product prior to filling
the vessel.  This assures fresh product and removes any air
or gas when dealing with liquids.  This can be done by
loosening the product purge valve a very small amount until
the product is purged.  After thorough purging, the
product purge valve should be tightened.

2c: The product pressure gauge reading should be 10-50 psi
below the pre-charge pressure gauge reading. By carefully
opening the pre-charge valve, the pressure becomes equal-
ized, then begins to drop below the product pressure.  The
pre-charge valve should be carefully controlled so as to not
vent the pre-charge gas too fast.

2d:  When the cylinder becomes a maximum of 80% full
(see volume indicator), all valves should be closed.  The
product connection is slowly broken in order to vent any
trapped product.  After vessel removal, all connections
should be checked for leaks and the pre-charge and product
valve ports capped to prevent leakage.

2e:  Pack the DuraSite in appropriate carrying case to meet
D.O.T. guideline, with D.O.T.
paperwork and transport to lab for analysis.

STEP 3: FOR DIRECT CONNECTION TO
SAMPLER.

3a:  Connect the sampler discharge port to the product inlet
port to the DuraSite using 1/8" stainless steel tubing.

3b:  (Gas sampling)  Connect the pre-charge port to the
DuraSite to the pipeline for pre-charge pressure (Proceed to
step 3d), or configured like the liquid sample application
below. (Step 3c)

3c:  (Light sampling)  Pre-charge the DuraSite as indi-
cated in Step 1, then install a pressure relief valve to the pre-
charge port and open the pre-charge valve on the DuraSite.
(The pressure relief valve should have a relief pressure
setting of  approximately
100 psi above line pressure.)

3d:  Open the product inlet valve of the DuraSite and the
purge valve on the sampler. Next open the purge valve on the
product end of the DuraSite and allow product to purge all
lines and connections out.

3e:  Close purge valves and begin sample cycle.

3f:  At the end of sample cycle, close product inlet valve on
the DuraSite and remove the DuraSite. Pack the DuraSite
in appropriate carrying case to meet D.O.T. guideline, with
D.O.T. paperwork and transport to lab for analysis.

Step 4:  (In The Lab)  Prior to analysis, the product should
be mixed.  This is accomplished simply and efficiently by
inverting the cylinder end-over-end,  causing the mixing ball
to fall through the product.  Approximately 10-12 trips of the
mixing ball through the product assures a homogenous
solution.

Step 5:  The regulated pre-charge gas should be recon-
nected to the pre-charge side of the cylinder.  The pre-
charge gas supply should remain open during analysis.

Step 6:  Purging a small amount of product from the vessel
removes unmixed product from the tee, relief device, gauge,
etc.  The unit can now be connected to a chromatograph
and the product analyzed.

Step 7:  After analyzing, the remainder of the product
should be dumped and the vessel properly cleaned.  Normal
cleaning can be accomplished by rinsing the product end
with a petroleum solvent and flushing with acetone.  If a
more thorough cleaning is required, the vessel should be
disassembled.

WARNING:  A portable sample vessel should never be
filled  to more than 80%.  This allows a 20% pre-charge
cushion to absorb thermal expansion of the product.
Shipping:  Extreme care should be taken when preparing a
vessel for shipment.  Both valves should be capped to
prevent possible leakage.  The vessel should be placed in
a snug-fitting, well-padded and durable case.  All applicable
DOT regulations should be adhered to.

Summary of Contents for DynaPak 2010N Series

Page 1: ...DynaPak2010N SystemSupport Manual ANSI 900 Version05302001ATEXRev...

Page 2: ...rantysetforthherein YZ sentireliabilityandthecustomer sexclusiveremedywillbereplacementoftheproduct s or atYZ soption YZ sreasonableefforttomaketheproductmeetthewarrantysetforthherein YZdisclaimsallot...

Page 3: ...ning installation operation contact your YZ representative or YZ Customer Service at 936 788 5593 2 System Components The primary components of the DynaPak 2010N Sampling System are illustrated here 1...

Page 4: ...sig using DPS 2 C Proportional to flow sampling in this mode of operation the Z 65 counter operates as a dividing counter The Z 65 counter periodically energizes a low power solenoid valve As in the o...

Page 5: ...ch could cause aerosols or significant pressure drops 4 2 The sampler should not be located within the defined meter tube region AGA 3 manual A The number of unobstructed straight pipe diameters upstr...

Page 6: ...CAUTION Incorrect operation of valves over tightening can result in damage to the valve components isolation valve bonnet assembly which might result in the valve stem being screwed out of the probe...

Page 7: ...high pressure supply valve without ensuring that the equalization valve is open Failure to do so may damage the DPS 2 s internal components d Open the other supply valve e Close the equalization valve...

Page 8: ...created 6 2 Variable volume constant pressure cylinder The free floating piston cylinder DuraSite may be installed in a horizontal position on an optional vessel rack Free floating piston cylinders sh...

Page 9: ...b on the top of the pump counterclockwise to set the pump displacement at 4cc stroke 7 4 Move all of the mode switches on the Z 65 to their off positions 7 5 Move both timer counter dials to the 0 pos...

Page 10: ...E The time 18 minutes above corresponds to the dial setting shown for the Z65 model with the timer range setting in the factory position jumper on the two left pins See section 12 4 Timer Range Settin...

Page 11: ...on 8 6 Turn mode switch 3 to on 8 7 Turn mode switch 4 to off 8 8 Press the test button once to initiate the timer sequence 8 3 Adjust the pump volume adjustment knob to the value used in the calculat...

Page 12: ...time 18 minutes above corresponds to the dial setting shown for the Z65 model with the timer range setting in the factory position jumper on the two left pins See section 12 4 Timer Range Setting 3 2...

Page 13: ...on 9 6 Turn mode switch 3 to off 9 7 Turn mode switch 4 to off 9 8 Press the test button once to initiate the timer sequence NOTE For the Z 65 to operate the DP 2010N System a minimum pressure differ...

Page 14: ...ng connections to the Z 65 terminal strip see illustration b Turn mode switch 1 to on c Turn mode switch 2 to off d Turn mode switch 3 to on e Turn mode switch 4 to off OR 10 3 If the input is a volta...

Page 15: ..._______ 5 pulses volume metered pulses MMCF or pulses MCM e __________ 6 counter setting a x c x d x e b Example 1 English Gas Flow Units Example 2 Metric Gas Flow Units pumpdisplacement a 2cc 2cc sam...

Page 16: ...e into the memory Sample pump displacement per stroke 1cc 2cc 4cc Number of turns open on the pump volume knob 3 6 12 NOTE If the calculated counter setting is less than 1 or greater than 99 the pulse...

Page 17: ...DynaPak 2010N Version 05302001 ATEX Rev YZ Systems Inc 3101 Pollok Drive Conroe Texas USA 77303 P 936 788 5593 F 936 788 5720 17...

Page 18: ...mp body d Screw the stroke adjustment knob all the way down to the 0 cc stroke setting e Unscrew the pump body by hand from the inlet check valve assembly Separation at this point is recommended to ma...

Page 19: ...all o rings and bushings to prevent damage j Install the body bushing into the bottom of the pump body NOTE Apply a light coat of assembly grease on the plunger shaft prior to installation k Insert a...

Page 20: ...te if the battery is low NOTE The solenoid must be connected to test the battery condition Battery condition cannot be tested with a volt meter Replacing a Depleted Battery 1 Remove the four thumb scr...

Page 21: ...s for the DynaPak 2000N Series gas samplers Part Number Description Qty Location C4 0004 Filter element 1 see diagrams 3 and 4 D3 0002 DP 2000N pump seal kit 1 see diagrams 1 and 2 D3 0003 YZ filter r...

Page 22: ...s sensing flow in the pipeline the green LED should illuminate If flow is not present the red LED should illuminate 12 3 LCD Stroke Indicator Test Mode A To test the stroke counter set the mode switch...

Page 23: ...e factory position far left position 12 4 Timer Range Setting A There are two Z 65 models the Z 65 6 1 and Z 65 6 03 Each Z 65 timer has two ranges for the timer setting dials 1 Z 65 6 1 Range Setting...

Page 24: ...minutes set the configuration jumper to the center position Note To obtain maximum battery life choose the longest solenoid stroke rate possible Mutual System Approved Factory factory positioned for 1...

Page 25: ...he green LED will turn on and off and the red LED will illuminate again This will normally occur very quickly and give the appearance that the green LED blinks on when the pulse input is received and...

Page 26: ...N B1 0020 Inlet Screen P N C4 0006 Inlet Check Valve Spring Guide Inlet Check Valve Spring O Ring 006 P N A5 1006 O Ring 012 P N A5 1108 O Ring 108 P N A1 0099 Male Connector 1 8 P X 1 8 T P N B1 0015...

Page 27: ...ing P N B1 0051 P N C3 0005 P N B1 0030 Set Screw P N C0 0096 P N C0 0014 Cap Screw 6 Each O Ring 012 P N B1 0020 P N C0 0026 Inlet Screen P N C4 0006 P N C3 0008 P N C4 0022 20 Micron Filter Inlet Ch...

Page 28: ...28 Diagram 3 YZ filter regulator assembled...

Page 29: ...rt A3 0073 O Ring 012 Retaining Nut P N A5 1012 Seat P N A3 0072 Upper Dart P N A3 0071 Upper Spring P N C3 0010 P N A3 0070 Piston P N A3 0069 Piston Spring Spring Cap P N A3 0068 P N A3 0067 Bonnet...

Page 30: ...less Steel Set Screw Inlet Rupture Disc 1800 psi P N A3 0083 Rupture Disc Retainer P N A3 0084 YZ Valve Body P N A3 0137 0 1000 PSI LinkPlus Model No C1 0005 0 2000 PSI LinkPlus Model No C1 0006 0 600...

Page 31: ...P N E0 5001 Magnet Diaphragm SS Hex Nut P N C0 0048 Qty 8 P N A6 0013 P N G2 0002 Cable Assembly 30 Ft Shrink Wrap SS Bolt P N C0 0047 Qty 8 P N H2 0002 Qty 2 Magnet Support Body Diaphragm Body Diaph...

Page 32: ...65 6 Controller Assembly 1 1 Model Z 65 6 1 F2 0001 Model Z 65 6 03 F2 0018 2 Battery Pack E3 2001 1 3 Stroke Counter Assembly G1 0001 1 4 Terminal Strip 6 Position H1 0001 1 5 BCD Switch E1 0001 2 6...

Page 33: ...on 05302001 ATEX Rev YZ Systems Inc 3101 Pollok Drive Conroe Texas USA 77303 P 936 788 5593 F 936 788 5720 33 Diagram 8 Z 65 Installation Notes Wiring Control Documentation System Installation Notes a...

Page 34: ...hould be closed The product connection is slowly broken in order to vent any trapped product After vessel removal all connections shouldbecheckedforleaksandthepre chargeandproduct valveportscappedtopr...

Page 35: ...t Head YZ UniValve Purge Valve Nut P N A3 0080 P N A3 0063 Valve Seat P N C6 1003 Ser 4608 P N 12 1006 Ser 0321 and Prior P N 12 1007 Ser 0322 0613 P N 12 1008 Ser 0614 4087 P N C6 1002 Ser 4088 Pisto...

Page 36: ...36 Notes...

Reviews: