background image

5

6

8.Measurement accuracy- in the absence of movement:

Note:The accuracy(A

RMS

) is calculated by the measurement values after a statistical distribution;

compared to the reference device in a control study, approximately 2-thirds of the values were at(over 
or below) the accuracy(A

RMS

) value.

Pulse rate: 25bpm~250bpm, accuracy(A

RMS

): ±1% or ±1bpm(larger)

9.Operation mode: Continuous operation.

65

70

75

80

85

90

95

100

0

15

30

45

60

75

90 105 120 135 150

Time�/s

Response

Time

8�

Beat

Average

SpO

2

/ %

Figure 4

10.Device response time.(See Figure 4)     
11.Peak wavelengths and light output power:
Emission wavelength range 600nm-1000nm, radiation intensity is less than 
15mW/sr (20mA).
Information of wavelength range may be of especial use to clinical doctors.
12.Description of the effect on displayed and transmitted SpO

2

 and Pulse rate 

data value:
1)data averaging and other signal processing.
2)the data update period:≤3 pulse rate cycles, less than 30s.
Note: Data processing and update will not affect pulse rate and SpO

2

.

13.The pulse waveform has been normalized, the measurement value is the best when the waveform is 
smooth and stable.
14. Internally Powered ME Equipment
15.TYPE BF APPLIED PARTS
16.Degrees of protection provided by enclosures (IP code):IP22.

Technical Description

1. Determination of Oxygen Saturation Accuracy 
The claimed oxygen saturation is supported by coverage of the entire range of clinical research 
measurements. 

2.Data Collection
In the clinical test process, data points are recorded with comparable density in the entire claimed 
range.
3. Data Analysis
For each claimed range, the oxygen saturation accuracy of the Pulse Oximeter should be represented 
in the form of mean root square of the difference between the measured values of oxygen saturation 
and the reference value. The formula is as follows:
A

rms

: accuracy. 

n: test sample quantity. 
SpO

2i

 : measured value of pulse oxygen saturation during the first measurement using the finger pulse 

Oximeter. 
S

Ri

: reference value of pulse oxygen saturation during the ith measurement using the carbon-monox-

ide-blood-gas analyzer.
4. Characteristics of Population under Clinical Research 
The group of healthy adult volunteers for the study consisted of 12 subjects - 4 women and 8 men, The 
ages ranged from 18 to 50 years. The subject weights ranged from 48 to 90kg. The subject Height 
ranged from 160 to 182cm. The Skin tones included in the study were as follows: 3 subjects with dark 

Indication for Use

Intended use: The YUWELL

®

 Finger Pulse Oximeter is a kind of non-invasive, non-sterile, reusable, spot 

checking device which can measure and display SpO

2

 and pulse rate through finger. It is intended for adults 

and children (weight >30 kg) and is expected for home and hospital inspection. The device is not for continu-
ous monitoring, using during motion or using with low perfusion.
Contraindications: None.

Technical Parameters

1.Display Type: LED 
SpO

2

 Display range: 0%~100%.

Pulse Rate Display range: 25bpm~250bpm.
2.Power: Two AAA 1.5V alkaline batteries.
3.Working Current: Less than 40mA at rated voltage 3V.
4.Anti-interference ability of ambient light: Deviation in blood oxygen content is less than ± 1% when 
measured under indoor nature light / existing lighting and measured in the dark room.
5.The product will automatically shut down when there is no signal detected for about eight seconds.
6.Dimension(YX102/YX103):60mm*38mm*35mm(LWH), Weight:38g approximately (without batteries).
7.Working Environments: Ambient temperature: 5℃~40℃; Relative humidity: ≤80%; Atmospheric 
pressure: 860hPa~1060hPa.

Display

Symbol of oxygen saturation

Low Battery

Symbol of 

pulse rate

Pulse bar

Pulse rate

Oxygen

saturation

Figure 2 Wave signal undetected

Figure 3

Signal undetected 

Signal inadequacy

 (eg: 1.Finger is out. 

2.Device fault )

Range

90% to 100%

80% to 90%

70% to 80%

<70%

A

RMS

±2%

No definition

pigmentation, 5 subject with very light pigmentation. 
The remaining subjects with light (medium) skin 
tones of Asian origins.
5.Graphical Plot of SaO

2

 versus error (SpO

2

 - SaO

2

)

6.The technology used in Finger Pulse Oximeter has 
been verified with accuracy when there is no motion 
via human blood studies on healthy volunteers of 
both male and female with light to dark pigmented 
skin in induced hypoxia studies in the range of 
70%-100% SpO

2

 against a laboratory co-oximeter.

7.The technology used in Finger Pulse Oximeter has 
been verified with the pulse rate accuracy of 
25-250bpm range in the bench top test against 

SpO

 ACCURACY (A

RMS

)

Summary of Contents for YX102

Page 1: ...ue to the limited size of the label the font is too small please put it at a suitable location for viewing All specifications and product configurations are subject to change without notification YUWE...

Page 2: ...ous oxide to avoid the risk of explosion 15 Caution When using this product you only need to fix it on your fingers Excessive pressure on your fingers may cause skin damage 16 Caution The maximum skin...

Page 3: ...ight can be focused onto human nail tip through perspective clamp finger type sensor Then measured signal can be obtained by a photosensitive element information acquired through which will be shown d...

Page 4: ...cy The claimed oxygen saturation is supported by coverage of the entire range of clinical research measurements Indication for Use Intended use The YUWELL Finger Pulse Oximeter is a kind of non invasi...

Page 5: ...monox ide blood gas analyzer 4 Characteristics of Population under Clinical Research The group of healthy adult volunteers for the study consisted of 12 subjects 4 women and 8 men The ages ranged fro...

Page 6: ...ell should be maintained from dust pollution use a soft cloth or lint free cleaning agent with the sponge to wipe Make sure no liquid will enter into the equipment Disinfecting The recommended disinfe...

Page 7: ...aration Radiated Emission Emissions test RF emissions CISPR 11 RF emissions CISPR 11 Compliance Group 1 Class B Table 1 For all ME EQUIPMENT and ME SYSTEMS RF emissions CISPR 11 Emissions test Result...

Page 8: ...resent actual modulation it would be worst case 1700 1990 2400 2570 GSM 1800 CDMA 1900 GSM 1900 DECT LTE Band 1 3 4 25 UMTS Bluetooth WLAN 802 11 b g n RFID 2450 LTE Band 7 1720 1845 1970 2450 Pulse m...

Reviews: