YSI ProDSS Use, Calibration And Maintenance Manual Download Page 18

YSI ProDSS Meter Calibration and Maintenance 

K:\Streamkeepers\Monitoring\Eqpt & supplies\Calibration-Maintenance\SOPs\YSI ProDSS maint-
cal SOP's  
 

18

 
Titration Steps: 
NOTE:

 dilute the chemicals going into the sink during the process with a continuous stream of 

tap water to prevent damage to plumbing. 
 
1.  Put on the plastic apron, face shield and Nitrile/Latex gloves. 
2.  Pour off the water seal and invert the bottle several times to mix the floc. 
3.  Remove the glass stoppers and put sample bottles in sink. 
 
4.  Using a pipette, draw and add 2 mL of sulfuric acid to each sample bottle. Keep the tap water    

running. 

5.  Rinse the pipette and put the cap securely on the acid bottle. 
6.  Put the stoppers back on the bottles and invert them several times over the sink until the 

precipitate has completely dissolved. 

7.  To transfer the liquid to the 203 mL volumetric flask, place the flask on top of BOD bottle 

number 1 and invert over sink. Once the flask is full, quickly remove the BOD bottle. The 
liquid should appear slightly convex on the top of the flask. Transfer the sample to the #1 
Erlenmeyer flask. Repeat for samples # 2 and 3. 

8.  Carefully uncover the sodium thiosulfate burette (a purple cover on it for protection and 

storage.) 

9.  Ensure there is enough thiosulfate solution to complete the three trials in the burette. Add 

solution to the reservoir if necessary. 

10. Empty any of the sodium thiosulfate left in the burette. 
11. Turn the stopcock valve so the handle faces the burette body (directly away from you toward 

the wall). 

12. Tighten the valve on the rubber bulb and pressurize the reservoir. Pump the rubber bulb gently 

to raise the solution, drain, refill and drain twice more to rinse, then fill the burette until the 
sodium thiosulfate escapes from the top nipple.  

13. Turn the stopcock 90 degrees and loosen the valve on the bulb to allow the excess solution to 

return to the reservoir. 

14. Drop a stir bar in the first sample flask and place it centered on the magnetic stirrer. Place the 

stirrer and flask centered under the burette valve to ensure drips fall cleanly into flask. 

15. Turn the stirrer on the lowest setting. 
16. Turn the stopcock valve handle toward you (there is a guide mark) and begin adding the 

sodium thiosulfate drops. You can add drops somewhat quickly at first but slow down when 
the sample starts turning a pale yellow. 

17.  Now, SLOWLY add one drop at a time until the sample turns very pale yellow and stop 

titrating. (It helps to place white paper between the burette and sample bottle.) 

18. Add 1 to 2 mL of the starch solution into the flask using the squirt bottle until the sample turns 

a dark purple. 

19. Continue to titrate the sample, adding drops VERY SLOWLY until the purple color just 

disappears and the sample is clear. Record the volume of solution used for each sample, to the 
nearest 0.1 mL. If ANY doubt about the end point, you can check the titration of the sample. 

20. Check the end point by adding a drop of potassium bi-iodate into the flask with a dropper that 

delivers 20 drops per mL, identical to the burette. If the end point is correct, the purple color 
should reappear. If more than one drop is required, then the end point was overrun. 

 
 

Summary of Contents for ProDSS

Page 1: ...Sensor Installation Sensor installation Calibration Barometer Conductivity pH Turbidity Dissolved oxygen Maintenance Short term storage Long term storage A FEW NOTES BEFORE GETTING STARTED The instru...

Page 2: ...and unit short term sensors are still attached to the unit and long term sensors have been removed and individually stored If the unit has been stored under the short term procedures continue to the C...

Page 3: ...y tighten the retaining nut if any resistance is felt loosen the nut and try again Finger tighten as much as you can then use the sensor installation tool to tighten until snug see image below DO NOT...

Page 4: ...d Once aligned push the sensor into the bulkhead until it is firmly seated in the port see image below 5 Carefully tighten the retaining nut if any resistance is felt loosen the nut and try again Fing...

Page 5: ...o finish charging When using the AC adapter it takes approximately 14 hours to charge the ProDSS battery when the instrument is turned off during the charge The amount of time required to completely c...

Page 6: ...b practices This includes what was calibrated who calibrated it and when it was calibrated To prevent contamination always calibrate the sensors in the following order Barometer if necessary conductiv...

Page 7: ...fore QC verified data can be collected in the field Calibration 1 If barometer units are not in inHg refer to Check Sensor Readings on page 4 to change units 2 Select the calibration button and select...

Page 8: ...it sit for about 30 seconds Use the cleaning brush from the maintenance kit to scrub the inside of the chamber insert in holes and scrub back and forth about 20 30 times Rinse the probe in clean tap...

Page 9: ...sor pH Calibration Equipment Needed pH Calibration Protocol Data Sheets notebook YSI ProDSS meter Purified water Kimwipes pH 4 7 and 10 buffer solutions A note about pH millivolts mV Millivolts mV are...

Page 10: ...ed meter Record results on the data sheet End of Season pH Calibration Check 1 Turn on and warm up instrument 2 Check and record readings of pH 7 and temp for the instrument in its kit pH 7 buffer Do...

Page 11: ...oint 2 you are now ready to calibrate with the next buffer solution DO NOT back out to the Run screen by hitting the left arrow button or the Esc button If you don t finish the three point calibration...

Page 12: ...ffer temperature reading value may be located on pH buffer bottle 7 Confirm stabilization by watching the actual readings on the screen if pH remains within 0 01 for 2 minutes the sensor is stable 8 A...

Page 13: ...e sensors and calibration cup with zero standard or deionized water Shake off the unit and cup after rinse 3 SLOWLY pour the deionized water down the side of the cup like you re pouring a beer fill th...

Page 14: ...in that range must be flagged an estimate Q in Streamkeepers flagging system and J or one of its variants in the WA Dept of Ecology s flagging system And if the meter does not perform to 1 0 or 10 of...

Page 15: ...d on the screen 8 Record results on the data sheet 9 Rinse the probe thoroughly with DI water before placing it back in the calibration cup this is important or it could mess up future calibrations 10...

Page 16: ...is essential to acceptance of our data by the State Equipment needed 1 or 2 large plastic bucket s depends on method used below DI water from lab DO Barometer Calibration Protocol Data Sheets notebook...

Page 17: ...the samples were taken on the data sheet WINKLER TITRATIONS IN THE LABORATORY Equipment for completing Winkler titrations Graduated cylinder burette with stopcock at bottom 25 mL with 3 way stopcock...

Page 18: ...y away from you toward the wall 12 Tighten the valve on the rubber bulb and pressurize the reservoir Pump the rubber bulb gently to raise the solution drain refill and drain twice more to rinse then f...

Page 19: ...tes and amounts of all solutions used to ensure there s enough unexpired liquids for the next calibration activity If not inform the office staff immediately Meter Samples 1 Place the bucket of prepar...

Page 20: ...on the ProDSS and its sensors follow the specified procedures in this section which have been taken directly out of the manual or it could void the warranty For more information on the warranty consu...

Page 21: ...f force 1 Remove the sensor from the bulkhead and soak the end of the sensor in a mild solution of clean tap water and dish soap in a glass beaker Soak for 10 to 15 minutes 2 After soaking rinse with...

Page 22: ...for installation Replacing the DO Cap NOTE Use extreme care to not damage the threads at the end of the sensor 1 Turn the old cap on the sensor counterclockwise and remove do not use any hand tools to...

Page 23: ...ly for correct transcription from the Calibration Code Label prior to selecting Update Sensor Cap Coefficients If you receive an error message after several entry attempts contact YSI Technical Suppor...

Page 24: ...ored more securely and under their own individual proper storing conditions Storage caps can be found in the individual storage boxes of each sensor do not lose these caps The following are a few guid...

Page 25: ...s only 4 Place the red cap onto the sensor connector end 5 Take out the storage bottle open lid and o ring from the storage box 6 Carefully slide the open lid with the o ring placed inside the lid ont...

Reviews: