22
23
S P E C I F I C A T I o N S
These specifications represent typical performance and are subject to change
without notice. For the latest product specification information, please visit
YSI’s website at www.ysi.com or contact YSI Tech Support.
Parameter
Range
Resolution
Accuracy
Temperature -5 to 55°C
0.1°C
± 0.2°C
Conductivity 0-500 uS/cm
0-5 mS/cm
0-50 mS/cm
0-200 mS/
cm (auto
ranging)
0.0001 to
0.1 mS/cm;
0.1 to 0 uS/
cm (range
dependent)
Instrument only: ± 0.5%
of the reading or 1 uS/
cm, whichever is greater.
Instrument with 1
or 4 meter cables:
± 1.0% of the reading or 1
uS/cm, whichever is greater.
Instrument with 10, 20,
or 30 meter cables:
± 2.0% of the reading or 1
uS/cm, whichever is greater.
Salinity
0 to 70 ppt
0.1 ppt
± 1.0% of the reading
or ± 0.1 ppt, whichever
is greater.
Total
Dissolved
Solids (TDS)
0 to 100 g/L.
TDS Constant
range: 0.3 to
1.00 (0.65
default)
0.0001 to 0.1
g/L (range
dependent)
Dependent on accuracy of
temperature, conductivity
and TDS Constant.
Summary of Contents for Pro30