59
9.10 Transducer surface temperature
EN 60601-2-37:2008+A1:2015 according to the requirements of clause 201.11, the
transducer surface temperature was measured under two conditions: transducer
suspension in the absence of air circulation environment (Still Air) and probe contact
phantom (Method B).
The maximum surface temperature and maximum temperature rise of the transducer
are listed below
:
Transducer
Test
method
Operation mode
Ambient
(
°C
)
Initial
temperature
(
°C
)
Final Radiating
Surface temp.(TF)
and/or Temp.Rise
(TR) (
°C
)
C5-2Fs
TMM
Normal(B+C+PW) 24.544
25.261
TF=30.020
TR=4.759
C5-2Fs
Still Air Normal(B+C+PW) 24.612
24.612
TF=35.528
TR=10.916
C5-2Ks
TMM
Normal(B+C+PW) 24.590
24.539
TF=29.08
TR=4.541
C5-2Ks
Still Air Normal(B+C+PW) 23.272
23.368
TF=32.995
TR=9.627
L11-4Ks
TMM
Normal(B+C+PW) 23.140
24.330
TF = 29.421
TR =5.091
L11-4Ks
Still Air Normal(B+C+PW) 22.380
23.340
TF = 34.73
TR =11.390
C8-5Ks
TMM
Normal(B+C+PW) 24.656
25.740
TF = 31.553
TR =5.813
C8-5Ks
Still Air Normal(B+C+PW) 23.731
23.838
TF = 38.376
TR =14.538
L11-4Gs
TMM
Normal(B+C+PW) 24.717
26.109
TF = 30.553
TR =4.444
L11-4Gs
Still Air Normal(B+C+PW) 23.827
24.037
TF = 35.407
TR =11.37
E10-4Ks
TMM
Normal(B+C+PW) 24.801
24.658
TF = 28.094
TR =3.436
E10-4Ks
Still Air Normal(B+C+PW) 24.747
24.275
TF = 33.943
TR =9.668