48
Medical ultrasonic coupling agent should be used in diagnosis.
CAREFUL
The probe might be damaged by repeated coupling agent over a
long period of time.
Check the probe housing frequently for cracks to avoid
damaging the internal components
,
The immersion of the
probe in water or liquid is forbidden .
7.4
Cleaning and disinfection
According to the intended use, the probe is in direct contact with the patient. In order to
avoid bacterial infection, after completing a test and turning off the equipment, please clean
and disinfect the probe according to the requirements of the instructions. Although the
handle does not directly contact the patient, it is an indirect transmission route of
pathogenic microorganisms, and the handle needs to be cleaned and disinfected if
necessary.
CAUTION
:
1. During cleaning or disinfection, please wear antibacterial gloves to
prevent infection.
2. After disinfection, please use sterilizing liquid to thoroughly remove the
residues. Chemical residue may be harmful to human body.
3. If you cannot confirm the efficacy of the disinfectant solution, contact
the manufacturer for product information.
⚫
Cleaning
WARNING
:
1.
Turn off the device after completing each examination, then clean
and disinfect the probes as required.
2 The probe is a non-immersion device and can never be submerged in
water.
3. It is forbidden to immerse the probe plug in a liquid such as water or
an anti - virus solution. Immersion can cause an electric shock or
malfunction.
4
If the coupling gel is not completely removed after each detection, it
will solidify and reduce the image quality of the probe.
5.
Do not place the probe in high temperature environment (over 55
℃
)
during cleaning and disinfection. High temperature may cause the
damage of the scanner..
6.
It is your responsibility to comply with the manufacturer's instructions
and your institution's policy on cleaning and disinfecting medical
equipment,Clean and disinfect your equipment in an appropriate
manner.