1
through one of these devices,turn off your analyzer.Before each usage,checking the status of your analyzer to ensure
it can operating normally.
Using short-wave diathermy,microwave diathermy,or therapeutic ultrasound diathermy(all now referred to as
diathermy)and electrocautery devices near this product may cause malfunction or lead to loss of essential
performance,please do not use analyzer near these equipment.Before each usage,observing the device to verify that
they are operating normally.
Guidance and manufacturer’s declaration-electromagnetic emissions
The device is intended for use in the electromagnetic environment specified below.The customer
or the user assure that it is used in such an environment.
Emissions test
Complianc
e
Electromagnetic environment-guidance
RF emissions CISPR11
Group 1
The device use RF energy only for its internal
function.Therefore,its RF emissions are very low
and are not likely to cause any interference in
nearby electronic equipment.
RF emissions CISPR11
Class B
The device is suitable for use in all
establishments other than domestic and those
directly connected to the public low-voltage power
supply network that supplies buildings used for
domestic purposes
Harmonic emissions
lEC61000-3-2
Not
applicable
Voltage fluctuations/
Ficker
emissions
lEC61000-3-3
Not
applicable
Guidance and manufacturer’s declaration—electromagnetic immunity
The device is intended for use in the electromagnetic environment specified below.The customer
or the user of the device should assure that it is used in such an environment.
Phenomenon
Basic
EMC
standard
or
test
method
Professional healthcare
facility environment
Home healthcare
facility environment
Electrostatic
discharge
IEC 61000-4-2
+/-8 kV contact
+/-2 kV,+/-4 kV,+/-8 kV,+/-15 kV air
Radiated RF
EM fields
IEC 61000-4-3
3V/m
80MHz-2.7GHz
80%AM at 1kHz or 2Hz
10V/m
80MHz-2.7GHz
80%AM at 1kHz
or 2Hz
Proximity
fields from RF
wireless
communications
equipment
IEC 61000-4-3
See the RF wireless communication equipment
table in"Recommended minimum separation distances".
Rated power
frequency
magnetic fields
IEC 61000-4-8
30A/m;50 Hz or 60Hz
Test specifications for ENCLOSURE PORT IMMUNITY to RF wireless communications
equipment
Test
Ban
Service
Modulatio
Maxi
Dista
Immunit