<5. Basic Operating Procedures>
49
IM 01E25A01-01EN
5.7 HART Configuration Tool
The connection of the HART configuration tool
(FieldMate (Versatile Device Management Wizard))
is shown as below. Read the user’s manual of HART
communication type as listed in Table 1.1 for the
detailed parameter setting via HART communication.
NOTE
• For more details regarding the operations of the
HART configuration tool, read the manual of HART
configuration tool.
• When using FieldMate, be sure that the revision is
R3.02.00 or later.
NOTE
Perameters on HART configuration tool are displayed
in English only. Even if any language other than
English is selected as “display language” from display
panel, parameters are displayed in English on HART
configuration tool.
5.7.1 Connections with HART Configura-
tion Tool
The HART configuration tool can interface with this
device from the control room, this device site, or any
other wiring termination point in the loop, provided
there is a minimum load resistance of 230 Ω between
the connection and the receiving instrument. To
communicate, it must be connected in parallel with this
device, and the connections must be non-polarized.
Read Figure 5.7.
F0537.ai
Iout+
Iout-
USB
FieldMate
Modem
USB
PC/FieldMate
Control Room
Magnetic Flowmeter
Terminal Box
Load
Resistance
230 to 600 Ω
Relaying
Terminals
4 to 20 mA DC
Signal
Transmission Line
Terminal
Board
Distributor
Figure 5.7 Connecting the HART Configuration Tool
IMPORTANT
Communication signal is superimposed on analog
output signal. It is recommended to set a low-pass filter
(approximately 0.1s) to the receiver in order to reduce
the output effect from communication signal. Before
online-communication, confirm that communication
signal does not give effect on the upper system.
5.7.2 HART Configuration Tool and
DeviceRevision
IMPORTANT
Protocol revision supported by HART configuration tool
must be the same protocol revision or later than that of
the device. If it is not, communication error occurs.
(1) Device Description (DD) and Device Revision
Before using the HART configuration tool, confirm that
the DD (Device Description) of this device is installed in
the configuration tool.
If correct DD is not installed to the configuration tool,
install a correct DD from the HART official site, otherwise,
contact the respective vendors of the configuration tool
for its upgrade information.
The device revision is as follows.
DD Revision
AXW4A: 1 or later
AXG1A: 1 or later
Device Type (transmitter)
AXW4A (0x371B)
AXG1A (0x371C)
Device Revision (Transmitter)
AXW4A: 1, 3
AXG1A: 2, 3
• Confirmation of DD revision
(a) Turn on the power of the configuration tool under
the standalone condition.
(b) Confirm the device revision from the installed DD
file name according to the procedure provided for
the configuration tool.
DD file name is four digits, upper two digits are device
revision and lower two digits are DD revision.
NOTE
Device revision of DD file is given in hexadecimal.
• Confirmation of Device revision
Connect the configurator to this device and confirm
the revision by the following parameter.
HART Communication Menu Path:
Device root menu ► Detailed setup ►
Device information ► HART setup ► Fld dev rev
Basic Operating Procedures