8-17
IM 01C50C05-01EN
Item
Parameter
name
Descriptions
Initial setting
Setup
Process
Sensor
→ Sensor1
Setup
Units
Sensor1 unit setting
degC
Lower Sensor
Limit/Upper
Sensor Limit
Maximum/minimum values
of Sensor1
Depend on
sensor setup
Minimum Span Minimum span of Sensor1
Damping Value Damping time constant in
seconds: 0.4 to 60.0s
*1
Probe Type
Sensor type setting.
*1
Probe
Connection
Single
N/A
Number of
Wires
Setting the number of
sensor connection wires
*1
Temperature
Standard
International Temperature
Scale of 1990 (ITS-90)
N/A
CJC Type
Selecting of Internal CJC or
Fixed CJC
Internal
sensor
Manual CJC
Temperature
Fixed CJC value setting
N/A
Manual CJC
Temperature
Units
Unit of Fixed CJC value
setting
N/A
Cable
Resistance
Value
Resistance of sensor cable
setting Set double of the
cable resistance between
the sensor and the device.
It is necessary for 2-wire,
3-wire and 4-wire RTDs/
ohms to compensate cable
resistance
5.0Ω
Process
Sensor
→ Electronics/
CJC Setup
Units
Unit of CJC setting
degC
Lower Sensor
Limit/Upper
Sensor Limit
Maximum/minimum values
of CJC
N/A
Minimum Span CJC minimum span
N/A
Damping Value Damping time constant in
seconds: 0.4 to 60.0s
0 .4
Cable
Resistance
Value
CJC cable resistance
0Ω