8-14
IM 01C50C05-01EN
8.4 Parameter List
Item
Parameter
name
Descriptions
Initial setting
Process variables
Digital Value
Device variable value in
engineering unit
(Sensor2, Average and
Difference is for future use)
N/A
Limit status
Data quality and limit status
of device variables
N/A
Data Quality
N/A
Loop current
Output value in mA
N/A
PV % rnge
Output value with respect to
the range in %
N/A
Dynamic variables
Digital value
(PV/SV/TV/
QV)
Dynamic variables
N/A
Limit status
Data quality and limit status
of PV/SV/TV/QV
N/A
Data Quality
N/A
Loop current
Output value in mA
N/A
PV % rnge
Output value with respect to
the range in %
N/A
Diag/
Service
Status
Sensor errors Show error status related to
a sensor. When any one of
the errors turns “ON”, check
the sensor and wiring
N/A
ADC errors
Show error status related to
analog to digital conversion.
When any one of the errors
turns “ON”, restart the
device. If error remains,
replace the device
N/A
Misc. errors
Show error status related to
a device. When any one of
the errors turns “ON”, restart
the device. If error remains,
replace the device.
N/A
Ext device
status
This bit is set if any Device
Variable is in an Alarm or
Warning State
N/A
Device
Diagnostic
Status 0
Device status of diagnostic N/A