6.2 Error Code Descriptions
and Corrective Actions
Information That Is Displayed When the Power Is Turned On (Display: Code.##)
Code
Description
Corrective Action
Refer To
3
You turned on the power while holding SET, so all the
settings have been initialized.
3.3
80
The system structure has changed, so all the settings
have been initialized.
―
87
The firmware of this instrument has changed, so all
the settings have been initialized.
―
Setting and Execution Errors (Display: Err.###)
Code
Description
Corrective Action
Refer To
759
Network initialization failed.
Check the network settings.
Chapter
4
1
812
The setting or command is not available during data
storage.
―
813
The setting is not available when an integration
operation is being carried out or aborted.
Reset the integration operation.
5.3
2
823
Cannot be changed during CAL.
Wait until CAL finishes.
8.3
2
832
You tried to load an empty setup parameter file.
Select a file that has had setup parameters saved
to it.
7.2
2
840
The setting or command is not available when auto
range is on.
841
You tried to start integration after the integration time
reached the specified timer time.
Reset the integration operation.
5.3
2
842
You tried to start integration during an integration
operation.
5.3
2
843
The integrated value overflowed during integration or
ended abnormally because of power failure or some
other reason.
Reset the integration operation.
5.3
2
844
You tried to stop integration when no integration
operation was taking place.
5.3
2
845
You tried to reset integration when the mode was not
integration mode or during integration.
5.3
2
846
You tried to start integration despite a peak
overrange.
―
847
When you tried to start integration in
continuous
integration mode, the integration timer time was set
to zero.
Set the integration timer correctly.
5.2
2
865
The setting is not available when an integration
operation is being carried out.
Abort or reset the integration operation.
5.3
2
886
Cannot be executed when the data update interval is
Auto.
Set the data update interval to other than Auto.
2.10
2
System Errors (Display: Err.###)
Code
Description
Corrective Action
Refer To
901
The setup data could not be backed up, and the
settings have been initialized.
The battery for backing up the setup data may have
worn out.
Servicing is required.
―
915
This is an EEPROM SUM error.
The EEPROM may be corrupted.
Servicing is required.
―
919
The currently installed modules and the setup data
conflict with each other. The setup data has been
initialized.
Servicing is required.
―
1 Communication Interface User’s Manual, IM WT310E-17EN
2 User’s Manual, IM WT310E-01EN