You can also connect to the circuit under measurement by placing a terminal block near the
WT332E/WT333E and connecting the power cables and voltage measurement cables to the
terminal block.
T
S
R
Terminal block
T
S
R
R
S
T
Wiring Procedure with the Three-Voltage, Three-Current Method
(3V3A)
You can configure the wiring for the three-voltage, three-current method (3V3A) by adding the following
wiring to the wiring for a three-phase, three-wire system.
Wiring to the Current Input Terminals
In the aforementioned steps 1 and 2, connect the current input terminals of element 2 to the
S-phase power cable.
Wiring to the Voltage Input Terminals
In the aforementioned steps 3 to 5, connect the voltage input terminals of element 2 so as to
measure the line voltage between phase R and phase T.
The result of wiring the three-phase, three-wire system shown in the figure above is for the three-
voltage, three-current method (3V3A) is shown below.
T
S
R
Terminal block
T
S
R
R
S
T
± V
± V
± V
R
S
R
T
T
S
SOURCE
LOAD
I1
±
C
±
I3 ±
C
U1
±
V
U3
V
R
N
S
T
U2
±
V
I2
±
C
2.9 Wiring the Circuit under Measurement for Direct Input