6-79
IM WT3001E-17EN
1
2
3
4
5
6
7
App
Index
Communication Commands
:MEASure:FREQuency?
Function
Queries all settings related to frequency
measurement.
Syntax
:MEASure:FREQuency?
Example
:MEASURE:FREQUENCY? ->
:MEASURE:FREQUENCY:ITEM1 U1;
ITEM2 I1
Description This command is invalid on models with the
frequency measurement add-on (/FQ) option,
because the frequency can be measured
simultaneously on all input elements.
:MEASure:FREQuency:ITEM<x>
Function
Sets the frequency measurement item or queries
the current setting.
Syntax
:MEASure:FREQuency:ITEM<x> {U<x>|
I<x>}
:MEASure:FREQuency:ITEM<x>?
<x> of ITEM<x> = 1 or 2 (Freq. 1 or Freq.2)
<x> of U<x>, I<x> = 1 to 4 (element)
Example
:MEASURE:FREQUENCY:ITEM1 U1
:MEASURE:FREQUENCY:ITEM1? ->
:MEASURE:FREQUENCY:ITEM1 U1
Description This command is invalid on models with the
frequency measurement add-on (/FQ) option,
because the frequency can be measured
simultaneously on all input elements.
:MEASure:FUNCtion<x>?
Function
Queries all settings related to user-defined
functions.
Syntax
:MEASure:FUNCtion<x>?
<x> = 1 to 20 (F1 to F20)
Example
:MEASURE:FUNCTION1? ->
:MEASURE:FUNCTION1:STATE 1;
EXPRESSION “UMN(E1)”;UNIT “V”
:MEASure:FUNCtion<x>:EXPRession
Function
Sets the equation of the user-defined function or
queries the current setting.
Syntax
:MEASure:FUNCtion<x>:EXPRession
{<String>}
:MEASure:FUNCtion<x>:EXPRession?
<x> = 1 to 20 (F1 to F20)
<String> = Up to 50 characters
Example
:MEASURE:FUNCTION1:
EXPRESSION “UMN(E1)”
:MEASURE:FUNCTION1:EXPRESSION? ->
:MEASURE:FUNCTION1:
EXPRESSION “UMN(E1)”
:MEASure:FUNCtion<x>[:STATe]
Function
Enables (ON) or Disables (OFF) the user-defined
function or queries the current setting.
Syntax
:MEASure:FUNCtion<x>
[:STATe] {<Boolean>}
:MEASure:FUNCtion<x>:STATe?
<x> = 1 to 20 (F1 to F20)
Example
:MEASURE:FUNCTION1:STATE ON
:MEASURE:FUNCTION1:STATE? ->
:MEASURE:FUNCTION1:STATE 1
:MEASure:FUNCtion<x>:UNIT
Function
Sets the unit to be added to the computation
result of the user-defined function or queries the
current setting.
Syntax
:MEASure:FUNCtion<x>:UNIT {<String>}
:MEASure:FUNCtion<x>:UNIT?
<x> = 1 to 20 (F1 to F20)
<String> = Up to 8 characters
Example
:MEASURE:FUNCTION1:UNIT “V”
:MEASURE:FUNCTION1:UNIT? ->
:MEASURE:FUNCTION1:UNIT “V”
Description This command does not affect the computation
result.
:MEASure:MHOLd
Function
Enables (ON) or Disables (OFF) MAX HOLD
function used in the user-defined function or
queries the current setting.
Syntax
:MEASure:MHOLd {<Boolean>}
:MEASure:MHOLd?
Example
:MEASURE:MHOLD ON
:MEASURE:MHOLD? -> :MEASURE:MHOLD 1
Description • The MAX HOLD operation starts when the
MAX HOLD function is specified by the user-
defined function and
:MEASure:MHOLd
is set
to ON.
• When :MEASure:MHOLd is set to OFF, the
MAX HOLD operation terminates, and the MAX
HOLD value becomes “no data.”
• If ON is specified while :MEASure:MHOLd is
ON, the MAX HOLD value is reset once, and
the MAX HOLD operation starts again.
• For details on the designation of the MAX
HOLD function, see the
User’s Manual IM
WT3001E-01EN
.
:MEASure:PC?
Function
Queries all settings related to the computation of
Pc (Corrected Power).
Syntax
:MEASure:PC?
Example
:MEASURE:PC? -> :MEASURE:PC:
IEC 1976;P1 0.5000;P2 0.5000
Description If the equation (
:MEASure:PC:IEC
) is set to
2011, “IEC 1993” is returned as a response.