GP-IB Interface
1.5 Setting the GP-IB Control
Procedure
START
STOP
REMOTE
CAL
MOTOR
SET
CURSOR
NULL
MENU
STORE SET
MENU
LOWER ITEM
LOWER FORM
ALL
USER SET
DISPLAY
ITEM & ELEMENT
RESET
SET
PAGE
PAGE
SCALING
HRM SET
MEASURE
SYNC SOURCE
AVG
LINE FILTER
FREQ
FILTER
FILE
MISC
STORE
IMAGE SAVE
HELP
NUMERIC
WAVE
OTHERS
ITEM
FORM
U / I / P
WP/q/
TIME
USER
ELEMENT
UPDATE
RATE
HOLD
SINGLE
INTEG
LOCAL
SHIFT
To exit the menu during operation, press
ESC
located above
the soft keys.
In the procedural explanation below, the phrase “press the
cursor keys
” may be used. This phrase refers to the
procedures for selecting items and entering values and
character strings. For details on the procedures, see section
3.14 in the
User’s Manual, IM WT3001E-01EN
.
Cursor keys
1. Press
MISC
to display the Misc menu.
2.
Press the
Remote Control
soft key to display the Remote Ctrl menu.
3.
Press the
Device
soft key to select GP-IB.
Only the communication interface selected here is enabled. This instrument does not
accept commands that are transmitted to other unselected communication interfaces.
4.
Press the
cursor keys
to set the address.
Explanation
Enter the following settings when using a controller to set information that can be
specified through key operation on this instrument or when outputting setup parameters
or output waveform display data to the controller.
Address
Set the address of this instrument within the following range for the addressable mode.
0 to 30
Each device that can be connected via GP-IB has a unique address within the GP-IB
system. This address is used to distinguish the device from others. Therefore, when you
connect this instrument to a PC, for example, make sure to assign a unique address to
this instrument.
Note
Do not change the address while the controller or other devices are using the GP-IB system.