![YOKOGAWA VZ20X Technical Information Download Page 51](http://html2.mh-extra.com/html/yokogawa/vz20x/vz20x_technical-information_3472729051.webp)
4.4.2 Data Monitoring Example
No.
Environment
Language
File name
1
Windows, Linux, etc.
Python
VZ20XSample07_EN.zip
2
FA
-M3
Ladder
VZ20XSample08_EN.zip
Refer to the sample programs when writing a program
to monitor measured values using function code 03.
The sample programs continuously read the CH1 to CH8 data of 40106 to
40121 from the VZ20X that is connected via Ethernet, by using a PC with
Modbus function code 03.
* See "Appendix" for details on preparing and running development
environments for sample programs.
The sample program can be downloaded from our website after entering
customer information.
URL
https://www.yokogawa.com/ns/vz/
4.4.2 Data Monitoring Example
Specify a register in which
the measured values of
VZ20X CH1 to CH8 are
continuous, and read out
the data.
Register number
(Reference no.)
Hex number
(hexadecimal)
Register name
R/W
Descriptions
40101
0064
Measured value time stamp
R
The time of measured values
can be read.
This setting is common to all
channels.
The coordinated universal time
or UTC can be read as a 64 -bit
unsigned integer.
40102
0065
R
40103
0066
R
40104
0067
R
40105
0068
Device status
R
The device status of the VZ20X
can be read.
40106
0069
CH1 measured value
R
Measured values can be read
as 32-bit signed integers.
The range changes depending
on the input range.
40107
006A
R
40108
006B
CH2 measured value
R
40109
006C
R
40110
006D
CH3 measured value
R
40111
006E
R
40112
0
06F
CH4 measured value
R
40113
0070
R
40114
0071
CH5 measured value
R
40115
0072
R
40116
0073
CH6 measured value
R
40117
0074
R
40118
0075
CH7 measured value
R
40119
0076
R
40120
0077
CH8 measured value
R
40121
0078
R
40122
0079
CH1 measured value status
R
The status of measured values
can be read.
40123
007A
CH2 measured value status
R
40124
007B
CH3 measured value status
R
40125
007C
CH4 measured value status
R
40126
007D
CH5 measured value status
R
40127
007E
CH6 measured value status
R
40128
0
07F
CH7 measured value status
R
40129
0080
CH8 measured value status
R
51
TI 77V01B01-11EN