<8. Error Message >
All Rights Reserved Copyright © 2006, Yokogawa Electric Corporation
IM 01C25R51-01E
8.
Error Message
Error message is shown in both pop up message and error log file (EJXMVTool.log).
F0801E.EPS
Figure 8.1 Error message example
In case of Physical Property error, code and sub code is shown in error log file
(EJXMVTool.log).
8.1
Error Message
Table 8.1 Error message
T0801E.EPS
Code
Sub code
Explanation
Recovery
Information Message
600
8
Rank deficient
The temperature or
pressure range which was
set by the user at the Fluid
Operating Range Setup is
too narrow. Try another
range value.
[Display]
The range specified in the step of
Fluid Operating Range Setup is too
narrow. Expand the range of Pressure
or Temperature.
[Log file]
Physical Calc lib Error - ERR_RANK