Standard Specifications 

IM 01G05B03-01E 1st Edition 

Fluid Temperature Range: 

General temperature type: 

–30 to +130°C (–22 to +266°F) 

High temperature type: 

–30 to +200°C (–22 to +392°F) 

Note:  Pay attention also to temperature specification of the 

couplant you choose. 

 

Connection Cable 

Connection cable is always necessary to connect the 
transducers (junction box) and the US300FM main unit. 
The length is specifiable from 1 m (3 ft) to 300 m (984 ft) 
by the meter. 

 

Accessories 

Fixing Hardware, Couplant, etc: 

Specified in the suffix code of US300FM/US300FT. 
Model codes as accessories also available for 
separate/additional orders (listed later). 

 

Data Transfer Software 

General: 

The software installed on a personal computer receives 
one or more records of logging data and parameter sets 
stored in US300FM main unit via RS232 
communication port. 
Data can be viewed or graphed on a PC monitor, or 
exported as a text file. 
This software works while US300FM is off-line (not 
measuring flow) where current output goes down to 
zero. 

Function: 

Displaying Parameter Record: 

Display parameter record of selected measuring data 
set. 

Displaying Measured Data in Table: 

Display measured data of selected measuring data 
set in table format. 

Displaying Data in Graph: 

Display measured data of selected measuring data 
set in graphic format. Marker type and color for each 
line of values selectable. Scales for time-axis and 
value-axis can be changed from default condition of 
automatic scaling. Graph printing function embedded. 

Displaying Statistical Data: 

Display statistical data of the measurement. Total 
data points, minimum, maximum, average and 
standard deviation of the measured data can be 
shown. Data range for statistical processing can be 
designated if necessary. 

Exporting Text File: 

Parameter record and measured data can be 
exported to a text file. Options for exporting items or 
their formats are available. 

Entering Remarks: 

User’s remarks for each measured data can be 
entered and edited in the transferred data file on a 
personal computer. Remarks can be displayed in the 
main window of the software. 

Display Language: 

English, German 

(Note)  Help is available only in English. 

Operating Environment: 

Personal Computer: 

Microsoft

®

 Windows

®

 hardware compatible, one or 

more RS232 port 

Operating System: 

Microsoft

®

 Windows

®

 98, ME, NT, 2000, XP 

Standard Accessories: 

RS232 cable, RS232 adapter 9/25 

* Microsoft

®

 and Windows

®

 are registered trademarks 

of Microsoft Corporation in the United States and/or 
other countries.

 

 

Units of Measurement 

Volume flow  Flow velocity  Mass flow

Totalizers 

Sound velocity 

 

 

 

Volume

Mass

 

m3/h m/s g/s 

m3 

g m/s 

m3/min inch/s  t/h  l 

kg 

 

m3/s  kg/h 

gal 

t  

l/h  

kg/min 

 

  

l/min    

 

  

l/s   

 

  

USgph     

   

USgpm     

   

USgps     

   

bbl/d    

 

  

bbl/h    

 

  

bbl/m     

   

1 gallon [US] = 3.78 l; 1 barrel = 42 gallons = 158.76 l 

Summary of Contents for US300FM

Page 1: ...ite White Perform soldering Finish the brown cables as well as white cables You do not connect the thin lines to any terminals Brown 12 2 Less than 10mm 12 2 Perform soldering Shield line Cable shield...

Page 2: ...ccurs A Tokuchu request is necessary for Submersible type However the diameter must be over 100 mm Changes to the section 3 5 5 Connection to the Power Supply page 20 AC power supply DC power supply T...

Page 3: ...y this behavior Addition and deletion to the section 4 1 1 Key Operations page 23 and related pages Refer to the following table for the function of corresponding key operations The correct explanatio...

Page 4: ...a wall thickness probe or some other ways and avoid mounting the transducers on such parts Attention Instructions for using Acoustic couplant weatherproof type Shin Etsu Silicone 1 COMPONENT RTV are a...

Page 5: ...the section 14 5 Activation of a Pulse Output page 95 Maximum pulse output rate is 2 pulse per second 20 Set the Pulse Value and the Pulse Width so that the actual maximum flow in the pipe is less th...

Page 6: ...Measuring Input One or two dual Channel A Channel B Current Output One or two Range 4 to 20 mA Load resistance 0 to 500 Note The current outputs may temporarily turn unstable during the power on seque...

Page 7: ...tion only for test service maintenance purposes Time Programmable Measurement Function Automatic start stop of the measurement without human operation using internal clock for specific systems Can be...

Page 8: ...suring data set in graphic format Marker type and color for each line of values selectable Scales for time axis and value axis can be changed from default condition of automatic scaling Graph printing...

Page 9: ...th 4 4 m 14 4 ft cable CH Large High with 4 4 m 14 4 ft cable DG Very large General with 12 m 39 4 ft cable N Always N Fixing band strap and clips 6 B For 25 to 1 400 mm 1 to 55 in Two fixing bands On...

Page 10: ...g fixture short ruler type for transducers medium pipe size type temperature range 30 to 200 C 22 to 392 F set of two blocks with 120 mm 4 7 in ruler USPA054 Mounting fixture standard type for transdu...

Page 11: ...flowmeter US300FM Unit mm inch Transducers US300FT B Transducers US300FT C US300FT D Transducers X m inch Y m inch X Y m inch Weight kg lb US300FT B 2 0 78 7 1 0 39 4 3 0 118 1 0 9 1 98 US300FT C 2 0...

Page 12: ...annel A P1 P1 Current output AR ARS Upstream transducer signal for channel A P2 P2 Current output when specifying two current outputs BV BVS Downstream transducer signal for channel B P3 P3 Frequency...

Reviews: