< 3. INSTALLATION, PIPING AND WIRING >
3-30
IM 12A0V2-E
Unit: mm
A
View A
600
500
530
Approx.
Approx.
200
Approx.
200
Maintenance
space
Maintenance
space
Maintenance
space
Front
For anchor bolts
4 - Ø15 holes
490
(20)
100
350
(100)
20
Liquid Analyzer for
Turbidity and Chlorine
FLXA402T
Pulse Generator
for Clean Unit *
1
PG400
Non-reagent Type
Free Available
Chlorine Sensor Unit
FC800D
Approx. 1320
530
550
150
50
80
93
PR1 : Pressure reducing valve
NV1 : Needle valve
BV1, BV2 : Ball valve
Weight : Approx.
45 kg
PR1
BV1
NV1
BV2
pH converter
KCI tank
Terminal box
pH sensor
pH holder
PH8EFP
PH8HF
VP40 pipe
Drain outlet
VP16 pipe
Sample inlet
FLXA402
Terminal box
cable mating connection
Ø6.5 to Ø12.5
*1: Attached when -P
is specified.
Figure 3.29
ST401G-PF7-
o
-A /R With free available chlorine analyzer, (ultrasonic oscillator), back
piping