< 3. INSTALLATION, PIPING AND WIRING >
3-14
IM 12A0V2-E
Unit: mm
100(
)
350
490
(20)
20
530
530
600
PUS400G (Note)
pH ultrasonic oscillator
(Cable O.D. 9 to 12 mm)
Terminal box
PR1
BV1
VP40 pipe
Drain outlet
250
150
150
Weight: Approx. 45 kg
(Note) Suffix code: -U (with ultrasonic oscillator)
VP16 pipe
Sample inlet
455
250
150
Approx. 1320
385
550
NV1
BV2
PR1: Pressure reducing valve
BV1, BV2: Ball valve
NV1: Needle valve
FC400G
PH8EFP
PH8HF
A
FC400G
FLXA402
Free available
chlorine converter
Free available
chlorine detector
100
Maintenance
Space
Approx. 200
Maintenance
Space
Approx. 200
Maintenance
Space
Front
Approx.500
View A
4-Ø15 holes
For anchor bolts
pH converter
KCl tank
pH holder
pH sensor
Figure 3.13
ST401G-PF6-
¨
-A With free available chlorine analyzer, pH converter, (ultrasonic
oscillator), (bottom piping)
Summary of Contents for ST401G
Page 1: ...User s Manual Model ST401G Sampling System IM 12A0V2 E IM 12A0V2 E 8th Edition ...