IM 12D08N04-01E
5-21
5. Parameter Setting
Code
Temperature compensation functions
20
21
22
23
24
25
26
27
28
29
Display
*T.R.˚C
*T.C.1
*T.C.2
*MATRX
*T1˚C
*T2..
*T3..
*T4..
*T5..
*L1xT1
*L1xT2
....
*L1xT5
*L2xT1
*L3xT1
*L4xT1
*L5xT1
Function
Set reference temp.
Set temp. coef. 1
Set temp. cpef. 2
Select matrix
Set temp. range
Enter conductivity
values for lowest
concentration
Concentration 2
Concentration 3
Concentration 4
Concentration 5
Function detail
Use >, , ENT keys to set value
Adjust compensation factor for mA1
output, if set to TC in section 5-2-5.
Set value with >, , ENT keys
Adjust compensation factor for mA2
output, if set to TC in section 5-2-5.
Set value with >, , ENT keys
Choose matrix if set to matrix comp.
in section 5-2-5, using >, , ENT keys
HCl (cation) pure water (0-80˚C)
Ammonia pure water (0-80˚C)
Morpholine pure water (0-80˚C)
HCl (1-5%, 0-60˚C)
NaOH (1-5%, 0-100˚C)
User programmable matrix
Enter 1st (lowest) matrix temp. value
Enter 2nd matrix temp. value
Enter 3rd matrix temp. value
Enter 4th matrix temp. value
Enter 5th (highest) matrix temp. value
Value for T1
Value for T2
....
Value for T5
Similar to code 24
Similar to code 24
Similar to code 24
Similar to code 24
Not used
X
1
2
3
4
5
9
Y Z Default values
25˚C
2.1%/˚C
2.1%/˚C