IM 12D08N04-01E
2-3
2. SC402G Specifications
2-2. Operating specifications
A.
Performance (under reference conditions with sensor simulation)
Converter output tolerance:
6
0.125% of output span
Conductivity (2
µ
S
3
K cm
-1
to 200 mS
3
K cm
-1
)
- Linearity
:
6
0.5%F.S.
- Repeatability
:
6
0.5%F.S.
Conductivity (1
µ
S
3
K cm
-1
to 2
µ
S
3
K cm
-1
)
- Linearity
:
6
1%F.S.
- Repeatability
:
6
1%F.S.
Resistivity (0.005k
Ω
/ K cm
-1
to 0.5M
Ω
/ K cm
-1
)
- Linearity
:
6
0.5%F.S.
- Repeatability
:
6
0.5%F.S.
Resistivity (0.5M
Ω
/ K cm
-1
to 1M
Ω
/ K cm
-1
)
- Linearity
:
6
1%F.S.
- Repeatability
:
6
1%F.S.
Note: "F.S." means maximum setting value of converter output. "K" means cell constant. YOKOGAWA
provides conductivity sensors which cell constant are 0.1 to 10 cm
-1
.
Temperature (Pt1000, PB36 NTC, Ni100)
- Linearity
:
6
0.3
°
C (190
°
C or more with Pt1000;
6
1
°
C)
- Repeatability
:
6
0.3
°
C (190
°
C or more with Pt1000;
6
1
°
C)
- Accuracy
:
6
0.3
°
C (190
°
C or more with Pt1000;
6
1
°
C)
Temperature (Pt100, 8.55k
Ω
NTC)
- Linearity
:
6
0.4
°
C
- Repeatability
:
6
0.4
°
C
- Accuracy
:
6
0.4
°
C
Temperature compensation
- NaCl table
:
6
1
°
C
- Matrix
:
6
3
°
C
Step response
:
90 % (< 2 decades) in 3 seconds
B.
Ambient operating temperature
: -10 to +55
°
C (10 to 130 ºF)
C.
Storage temperature
: -30 to +70
°
C (-20 to 160 ºF)
D.
Humidity
: 10 to 90% RH non-condensing