IM 12D7B2-E-H
7. SERVICE MODE
7-1. Introduction
Generally speaking their is no necessity to
adjust the settings of the service section All
parameters are pre-programmed to values
(so-called defaults) enabling you to start
working immediately.
The advanced functions available through
this section are only needed in some speci-
fic applications. This fine-tuning of the in-
strument gives a superior performance over
conventional 2-wire instruments.
If a function has to be adjusted it is called up
with the code mentioned. Having selected
the code then give you the possibility to
either activate or adjust the values for this
function. After this you will return to the
entry point to make other adjustments or go
back to the measuring status.
If errors are made during the programming
process, these will be indicated, no action
will be taken and you can start the program-
ming again.
28
Code
Routine
Use
Chapter
01
Temperature sensors and units
Select sensor type
7-3
Select °C or °F
02
Reference temperature
Adjust value
7-4
03
Output table or linear
Select which to use
7-5
04
Program output table
Set values in table
7-6
05
Cell constant
Adjust cell-constant
7-7 +10-1
06
Auto return function
Switch on or off
7-8
07
Measuring principle
Select 2 or 4 electrode
7-9
08
Temperature adjust
Correct cable error
7-10
09
Signal errors or output
Switch on or off
7-11
10
Temperature compensation factor
Adjust factor
7-12
11
%-indication on second
Switch on or off and
7-13
display line
Set values
12
Polarisation check
Switch on or off
7-14
13
Matrix compensation
Select 5x5 matrix compensation
10-2
14
Matrix temperature
Set temperatures for user matrix
10-3
compensation range
15-19
Matrix data
Set data for user matrix
10-4
20
“Soft”/“hard” fail situation
Select “soft” or “hard” fail
10-5
message per fail code
21
E6-fail message
Switch on or off
10-6
33
Passcode activation
Protecting levels
7-15
55
Restore defaults
Erase values
7-16