![YOKOGAWA SC200 Instruction Manual Download Page 28](http://html1.mh-extra.com/html/yokogawa/sc200/sc200_instruction-manual_3784901028.webp)
ESCAPE TO MEASURE can be used at any stage to abort operation. WARNING: If the HOLD function is activated the instrument returns with the question HOLD (flashing);
answer YES or NO or MODE again to return to measurement.
IM 12D7B2-E-H
24
5-3. Use of HOLD function
HOLD = HOLD output function
Display will blink HOLD and YES/NO
1. Access hold
2. Switch hold on/off
M E A S U R E
C A L
D I S P L A Y.
H O L D
YES
NO
YES
NO
MODE
MODE
M E A S U R E
C A L
D I S P L A Y.
H O L D
YES
NO
YES
NO
MODE
MODE
HOLD
Access maintenance mode
Select HOLD function
Move pointer to HOLD
Select HOLD function
Select HOLD
NO
NO
Switch hold OFF
Switch hold ON
NO
YES
MODE
NOTE: This function can only be used if activated
during commissioning (see chapter 4-3).
HOLD in left top of display is switched