6-6. Error messages and explanation
Code
Error description
Possible cause
Suggested remedy
E1*
Faulty signal from conductivity cell
Polarisation
Check cell is correct for range
Cell is fouled
Clean cell
E2
Wrong temperature coefficient during
Incorrect data entered
Repeat temperature calibration
calibration
E3
Calibration out of limits
Calibration value deviates more than
Check cell, connections or reference
20%
E5*
Conductivity too high
Measurement is out of range
Check cell range, reprogram range
E6*
Conductivity too low
Measurement is out of range
Check cell range, reprogram range
E7*
Temperature too high
Wrong temperature sensor selected
Program sensor type
Fault in connections
Check connections
E8*
Temperature too low
Wrong temperature sensor selected
Program sensor type
Fault in connections
Check connections
E10*
EEPROM write failure
Fault in electronics
Call yourYOKOGAWA service
organisation
E15
Temperature correction
Fault in cable or wrong
Check connections or program
outside limits (> 5
Ω
)
temperature sensor selected
correct sensor type
E17
Output span too small
0% range value set too high
Program value <60% of high range
E18
Output table makes no sense
Table changes from ascending to
Check programming of table
descending or vice versa
E19
Programmed value not accepted
Values not acceptable
Reprogram with new values
E20*
Data lost
Data, including initial setting, lost
Call your YOKOGAWA service
organisation
* These errors will trigger the FAIL-status.