Press the Back button to return to the ‘Test QIC’ menu
This test can be executed, independent from the current settings of the instrument.
-
Press setup button
-
Go to
Calibration
-
Select
Air Calibration
-
Select
Adjust
now
-
Press
Back
button
-
Go to
Commissioning
-
Go to
Advanced
setup
-
Go to
Test QIC
-
Set
Key
to ‘Factory Mode’
-
Set
SC Input
to the range in which you want to test the accuracy (see table below)
Set BOX 2 to the values listed below to simulate the resistivity (
Ω
) input.
Note:
In the QIC form a table is shown containig 27 fields.
For a proper accuracy test however a subset of five fields is recommended.
QIS 12D7B4-E-H
47
Range
Input
Display
Tolerance
2
100
Ω
100.0
Ω
±
0. 5
Ω
3
1
k
Ω
1.000 k
Ω
±
0.005 k
Ω
4
10 k
Ω
10.00 k
Ω
±
0.05 k
Ω
5
100 k
Ω
100.0 k
Ω
±
0.5 k
Ω
5
1
M
Ω
1.000 M
Ω
±
0.005 M
Ω
P t 1 0 0 0
Ω
RTD
P t 1 0 0
Ω
RTD
Resistance
Temperature
Tolerance
Resistance
Temperature
Tolerance
960.9
Ω
-10
°
C
±
0.3
°
C
96.1
Ω
-10
°
C
±
0.4
°
C
1097.4
Ω
25
°
C
±
0.3
°
C
109.7
Ω
25
°
C
±
0.4
°
C
1290.0
Ω
75
°
C
±
0.3
°
C
129.0
Ω
75
°
C
±
0.4
°
C
1460.6
Ω
120
°
C
±
0.3
°
C
146.1
Ω
120
°
C
±
0.4
°
C
1721.6
Ω
190
°
C
±
0.3
°
C
172.2
Ω
190
°
C
±
0.4
°
C
1904.6
Ω
240
°
C
±
1
°
C
Ni100
Ω
RTD
8 k 5 5
Resistance
Temperature
Tolerance
Resistance
Temperature
Tolerance
94.6
Ω
-10
°
C
±
0.3
°
C
47000
Ω
-10
°
C
±
0.4
°
C
114.2
Ω
25
°
C
±
0.3
°
C
8550.0
Ω
25
°
C
±
0.4
°
C
144.9
Ω
75
°
C
±
0.3
°
C
1263.0
Ω
75
°
C
±
0.4
°
C
175.9
Ω
120
°
C
±
0.3
°
C
343.0
Ω
120
°
C
±
0.4
°
C
231.8
Ω
190
°
C
±
0.3
°
C
4. Accuracy Test - Resistivity Input
5. Accuracy Test - Temperature Input
In the ‘Test QIC’ menu, set
Temperature
to the temperature element of which
you want to test the accuracy. Next vary the settings on BOX 1 as listed below
and check the readings on the display.
8 k 5 5
Resistance
Temperature
Tolerance
47000
Ω
-10
°
C
±
0.4
°
C
8550.0
Ω
25
°
C
±
0.4
°
C
1263.0
Ω
75
°
C
±
0.4
°
C
343.0
Ω
120
°
C
±
0.4
°
C