Connecting cable installation
General Instruction Manual
Wiring
IM 01U10B00-00EN-R, 3rd edition, 2018-07-09
45 / 90
7.3 Connecting cable installation
With remote type flow meters, sensors and transmitters are connected by means of con-
necting cables.
CAUTION
Risk of damage to the flow meter due to incorrect sealing
In case of metric cable entry ensure appropriate IP rating and suitability of O-ring of used
accessory (e.g. cable glands).
In case of NPT cable entry ensure appropriate sealing measures (e.g. use of sealing
tape).
In order to obtain optimum measuring results and ensure compliance with the specifica-
tion, it is imperative that an original connecting cable and original glands from Rota Yoko-
gawa are used. In order to ensure the IP code, the cable must be professionally installed
at the entries. If necessary, the cable may be shortened using the enclosed termination
kit. Refer to the cable termination instructions enclose to each termination kit that is at-
tached to each cable.
1
2
3
4
7
8
6
5
Fig. 29:
Cable gland parts mounting
1
Cap nut
5
Outer cable shield
2
Connecting cable
6
Outer cone part
3
Plastic part
7
Mounting thread
4
Inner cone part
8
Housing cable entry
If the connecting cable, included in the delivery, is too short, additional lengths can be
procured through the Yokogawa sales organization.