Installation
Installation instructions
26 / 90
IM 01U10B00-00EN-R, 3rd edition, 2018-07-09
▶
For Cleaning-In-Place (CIP) application, the standard minimum flow velocity of
1.5 m/s shall be used for cleaning the sensor. Volumetric flow shall be determined by
using the cross-sectional area at process connection.
Notes on fittings
and gasket
▶
General note: It must be ensured that the inner diameter (ID) of adjacent pipe
matches the ID of process connection of the sensor fitting to ensure 3-A compliance.
▶
For compliance of DIN 11851 process connection (process connection HS2) with 3-A
standards, a special sanitary gasket such as the k-flex gasket system by Kieselmann
GmbH, or similar must be used to retain 3-A compliance.
6.2 Installation instructions
Note the following rules prior to installation:
▶
Check packaging and contents for damage.
▶
Do not remove protective materials such as protective stickers or caps on process
connections until the start of the installation process.
▶
Dispose packaging materials in compliance with country-specific regulations.