IM 12B03D01-01E
6-6
6. Maintenance
6.7 Storage and O-ring/Gasket Replacement
• Storage Precautions
Care is required when storing the meter and sensor. To maintain in good condition,
observe the following:
(1) Before storage, wash off remaining sample solution from the sensor with water.
Deposits on the liquid junction, if any, must be removed thoroughly, or the
junction may be clogged. A clogged junction will cause invalid measurements,
e.g., unstable pH readings.
Keep a wetting cap attached to the sensor. For KCl refillable type sensors, seal the
fill hole with a plug.
(2) Leave the sensor connected to the meter body to protect the connectors and O-ring
from staining. Contamination may cause deteriorated insulation of connectors or
poor water resistance by the O-ring.
(3) Do not place any object on top of the sensor or on the top of the PH71 meter.
F0604.EPS
Wetting cap
Protective cover
Attach wetting cap containing cotton
wad moistened with a few drops of
water.
For KCl refillable type sensor,
seal fill hole with plug.
Figure 6.4
Sensor Storage Condition
• Storage Location
When not in use, store the meter and sensor in a safe place. If it is to be stored for a long
period, store it in a place:
• With low humidity at or near normal temperatures
• Not exposed to direct sunlight or water
• No corrosive gases are present
• Replacing the O-ring and Gasket
A gasket in the battery box and an O-ring on the sensor connector of the meter can be
replaced. Heavily contaminated or damaged gasket and/or O-ring should be replaced.
CAUTION
When installing an O-ring and gasket, clean them and their mounting surfaces with a
cloth moistened with alcohol so that they are free from dirt. Otherwise, water resistance
may not be assured.
Summary of Contents for PH71
Page 1: ...User s Manual IM 12B03D01 01E Model PH71 Personal pH Meter IM 12B03D01 01E 3rd Edition ...
Page 7: ...IM 12B03D01 01E 6 Contents ...
Page 17: ...IM 12B03D01 01E 1 10 1 Outline ...
Page 23: ...IM 12B03D01 01E 2 6 2 Preparation ...
Page 35: ...IM 12B03D01 01E 4 8 4 Calibration ...
Page 49: ...IM 12B03D01 01E 5 14 5 Keypad and Display Functions ...
Page 57: ...IM 12B03D01 01E 6 8 6 Maintenance ...
Page 73: ...IM 12B03D01 01E 8 10 8 Technical Information ...
Page 79: ...IM 12B03D01 01E 6 Appendix ...
Page 81: ...IM 12B03D01 01E MSDS 2 MSDS ...
Page 82: ...IM 12B03D01 01E MSDS 3 MSDS ...
Page 83: ...IM 12B03D01 01E MSDS 4 MSDS ...
Page 84: ...IM 12B03D01 01E MSDS 5 MSDS ...
Page 85: ...IM 12B03D01 01E MSDS 6 MSDS ...
Page 86: ...IM 12B03D01 01E MSDS 7 MSDS ...
Page 87: ...IM 12B03D01 01E MSDS 8 MSDS ...
Page 88: ...IM 12B03D01 01E MSDS 9 MSDS ...
Page 89: ...IM 12B03D01 01E MSDS 10 MSDS ...
Page 90: ...IM 12B03D01 01E MSDS 11 MSDS ...
Page 91: ...IM 12B03D01 01E MSDS 12 MSDS ...
Page 92: ...IM 12B03D01 01E MSDS 13 MSDS ...
Page 93: ...IM 12B03D01 01E MSDS 14 MSDS ...
Page 94: ...IM 12B03D01 01E MSDS 15 MSDS ...
Page 95: ...IM 12B03D01 01E MSDS 16 MSDS ...