41
IM 04P01B01-02E
Changing the Chart Speed
1.
Hold down the
MENU
key for 3 seconds to enter Setting mode.
2.
Press the
key to show
Chart
and then press the
key.
Set=Chart
Chart speed
Displays a description of the setup item.
3.
Set the chart speed and press the
key.
mm/h= 20
1/1500mm/h
Current chart speed.
Displays the range that can be specified.
(Display example on the dot model)
On the pen model, press the
key to select the chart speed.
5 6 8 9 10 12 15 16 18 20
24 25 30 32 36 40 45 48 50 54
60 64 72 75 80 90 96
100
120 125
135 150 160 180 200 225 240 250 270 300
320 360 375 400 450 480 500 540 600 675
720 750 800 900 960
1000
1080
1200
1350 1440
1500 1600 1800 2000 2160 2250 2400 2700 2880 3000
3600 4000 4320 4500 4800 5400 6000 7200 8000 9000
10800 12000
Chart speed on the pen model (unit: mm/h)
On the dot model, enter a vale to set the chart speed. The chart speed can be
set in the range of 1 to 1500 mm/h in 1 mm steps.
Press the
key to select the desired digit.
Press the
key to select the value.
4.
When the
Setting complete
screen appears, the new chart speed is applied.
Chart speed
Setting complete
Press the
ESC
key to change the chart speed again.
Hold down the
MENU
key for 3 seconds to return to Operation mode.
Viewing the Recorded Results
Pull the front cover tab of the chart cassette to open the front cover. The recorded chart
paper can be pulled out for viewing.
Front cover tab
Recorded chart paper can
be pulled out.
Recording/Displaying Data
* When the key or key is pressed while holding down the
SHIFT
key, the operation is
reversed as when the respective key is pressed by itself.