34
IM 04P01B01-02E
Setting the Input Range and Alarm on Measurement Channels
Selecting the Input Type
5.
Press the
key to select
1-5V
and then press the
key (see “Explanation”
on page 37).
Mode=1-5V
Scales and record
Setting Span Left
6.
Set Span_L to
1.000
and press the
key.
Press the
key to select the desired digit.
Press the
key to select the value.
Span_L= 1.000
0.800/ 1.200V
Displays the range of Span_L.
Setting Span Right
7.
Likewise, set Span_R to
5.000
and press the
key.
Span_R= 5.000
4.800/ 5.200V
Displays the range of Span_R.
Setting the Decimal Position and Scaling Left
8.
Display Scale_L.
Scale_L= 0.00
-200.00/ 300.00
Displays the scaling range.
9.
Press the
key to select the desired digit.
Scale_L= 0.00
-200.00/ 300.00
10.
Press the
key to select space and then press the
key (Scale_L is set to
0.0
).
Scale_L= 0.0
-200.00/ 300.00
Setting Scaling Right
11.
Likewise, set Scale_R to
500.0
and press the
key.
Scale_R= 500.0
-2000.0/ 3000.0
Displays the scaling range.
The
Setting complete
screen is displayed. When this screen is displayed, the
settings entered up to then are applied.
Finishing the Settings
12.
When
Setting complete
screen is displayed, press the
ESC
key.
The
Set=Range
screen is displayed.
03-03 Channel
Setting complete
Set=Range
Input range and
Setting the Unit
13.
Press the
key to select
Unit
and then press the
key.
Set=Unit
Engineering unit
Displays the description of setting item.
14.
Press the
key to set the first channel to
03
and then press the
key.
First channel Last channel
CH=03-03
First channel 01-
Displays the selectable range of channels.
15.
Likewise, set the last channel to
03
and then press the
key.
* When the key or key is pressed while holding down the
SHIFT
key, the operation is
reversed as when the respective key is pressed by itself.