<9. CALIBRATION>
9-6
IM 11P03A03-01E
6th Edition : Jan. 26, 2022-00
• If the sum of the zero calibration time, span calibration time and purge time is longer than
the calibration interval, the next calibration will be canceled. The internal timer works even if
the power supply to the instrument is interrupted while the automatic calibration function is
enabled. After power is restored, calibrations will be executed on schedule; there is no delay
due to the power interruption. Note that the internal timer does not work when the power
supply is interrupted during the calibration time.
• Once the automatic calibration function is turned off by setting in Function No. 57, the timer
will be reset.
• It is recommended that after the settings for the automatic calibration have been completed
manual calibration should be conducted once to verify whether the calibration is conducted
properly.