<2. SPECIFICATIONS>
2-2
IM 11P03A03-01E
6th Edition : Jan. 26, 2022-00
Power supply:
Power supply Voltage 100 to 115 V AC;
Rated voltage range: 100 to 115 V AC
Allowable voltage range: 90 to 127 V AC
Reted frequency:
50 or 60 Hz
Allowable frequency range: 48 to 63 Hz
Power supply Voltage 200 to 240 V AC;
Rated voltage range: 200 to 240 V AC
Allowable voltage range: 180 to 264 V AC
Reted frequency:
50 or 60 Hz
Allowable frequency range: 48 to 63 Hz
Power consumption:
100 to 115 V AC;
Max. 110 VA, normaly approx. 25 VA
200 to 240 V AC;
Max. 125 VA, normaly approx. 35 VA
KC Marking: Korea Electromagnetic Conformity Standard
GB:
GB30439 Part 1
Installation altitude:
2000 m or less
Installation category: II
Pollution degree:
2
Note: Installation category, called overvoltage category,
specifies impulse withstand voltage. Pollution degree indicates the degree of
existence of solid, liquid, gas or other inclusions which may reduce dielectric strength.
Materials in contact with gas: SUS316 stainless steel, Fluorine-contained rubber
Line connection:
Rc1/4
Conduit connection port: Ø27 hole
Installation:
Indoor, panel or wall mounting
Structure:
General purpose
Dimension:
406 (W) x 288 (H) x 216 (D) mm
Weight:
Approx. 18 kg
l
Characteristics
Repeatability: ±1% or less of F.S.
Linearity:
±1% or less of F.S.
Response time: 90% response within 3 sec. (from changing analog output at measured gas flow
rate 200 ml / min.)
Zero drift:
±1.5% or less of F.S. / Week
Span drift:
±2% or less of F.S. / Week
Temperature drift:
±1.5% or less of F.S. / 10 °C
Effects of measured gas flow rate:
±1% or less of F.S. for the rated flowrate ±10%