<7. START-UP>
7-4
IM 11P03A03-01E
6th Edition : Jan. 26, 2022-00
(5) Press the SPAN key. Turn off the zero gas flow and apply the span gas at the specified flow
rate. When solenoid valves for automatic calibration are installed, pressing the SPAN key
closes the solenoid valve in zero gas line and opens the solenoid valve in span gas line.
(6) The oxygen concentration on the DATA display changes. When the reading has stabilized
completely, press the EXEC key. The instrument is now calibrated to that value as the span
point.
(7) Turn off the span gas flow. Apply the sample gas and switch to MEAS mode.