8.  MAINTENANCE 

The product starts running immediately when the power is turned 

on; however, it needs 10 to 15 minutes of warm-up before it 

meets the specified performance. 

For cleaning the instrument, use a soft and dry cloth.

8.1  Calibration Apparatus 

 •

A calibrator (YOKOGAWA CA500 or equivalent)

 • Setting tool for adjustment (Refer to "4.2 Connector for 

Communication" in this manual.)

8.2  Calibration Procedure 

1. 

Connect the instruments as shown in figure below. First adjust 

the output-1 signal and then the output-2 signal.

Input

R: Externally connected 

     for current input

Shunt resistor

(Externally connected 

for current input)

Calibrator

Calibrator

Calibrator

Calibrator

Calibrator

Output-1

Alarm-1

Alarm-2

Output-2

10

11

3 2 1

4

5

6

7

8

9

Power supply

11

10

8

L

+

N

GND

1

3

+

2

5

+

2

5

ALM1

(N.O.)

COM

+

6

5

ALM2

(N.O.)

COM

+

7

9

+

2.  Use the calibrator and apply input signals equivalent to 0, 25, 

50, 75, and 100% of the input span to the product.

3. 

Then, check that the converter’s output signal shows a 

frequency corresponding to 0, 25, 50, 75, and 100% of the 

input span within the rated accuracy range. 

 

Take the same calibration procedure for the output-2, and 

confirm that the frequency of the output 2 is the same as that 

of the output 1.

For alarm output, check the relay action by the alarm indicator 

lamp or resistance of output terminals.

 • When adjusting the input signal, use the VJ77 parameter 

setting tool.

For adjustment using a setting tool, refer to the User’s Manual for 

setting tool and “8. LIST OF PARAMETERS” in this manual.

User’s Manual for VJ77 [Document No.: IM 77J01J77-01E]

9.  SAFETY AND EMC STANDARDS

The following will be acquired.

Safety:

IEC/EN 61010-1 compliance (CE)

IEC/EN 61010-2-030 compliance (CE)

CAN/CSA-C22.2 No.61010-1 (CSA)

CAN/CSA-C22.2 No.61010-2-030 (CSA)

UL61010-1 (CSA NRTL/C)

UL61010-2-030 (CSA NRTL/C)

Overvoltage category: II

Pollution degree: 2

Measurement category: O (other)

Rated power supply voltage:  15-30VDC( ± 10% ) 2.3 W 

100-240VAC/DC( ± 10% ) 

50/60 Hz 2.3 W 6.0 VA 

Rated voltage (Input signal) : ±10 V DC

Rated voltage (Alarm output signal) : 30 V DC  1A

Rated voltage (Communication output signal) : 5 V DC

Rated voltage (

Voltage Pulse output signal

) : 26.4 V DC

Rated transient overvoltage: 1500 V (*)

* This is a reference safety standard value for the 

measurement Category O. This value is not necessarily 

a guarantee of instrument performance.

 CAUTION

This instrument is for Measurement Category O 

(other). Do not use it for measurements in locations 

falling under Measurement Categories II, III, and IV.

    

Internal Wiring

Outlet

Entrance 

Cable

III

T

O

II

IV

Measurement 

category

Description

Remarks

O

(other) For measurements 

performed on circuits not 

directly connected to MAINS.

II

CAT.II For measurements 

performed on circuits directly 

connected to the low-voltage 

installation.

Appliances, 

portable 

equipments, 

etc.

III

CAT.III For measurements 

performed in the building 

installation.

Distribution 

board, circuit 

breaker, etc.

IV

CAT.IV For measurements 

performed at the source of 

the low-voltage installation.

Overhead 

wire, cable 

systems, etc.

EMC standards:

• CE:

EN 61326-1 Class A, Table 2

EN 61326-2-3

* The instrument continues to operate at a measurement 

accuracy of within ±20% of the range during testing.

EN 55011 Class A, Group 1

EN 61000-3-2

EN 61000-3-3

• RCM: 

EN 55011 Class A, Group 1

KC:

Electromagnetic wave interference prevention standard, 

electromagnetic wave protection standard

 CAUTION

Caution to comply with EMC standards: When 

operating this instrument by external power supply, 

use an independent power unit conforming to CE-

marking. Be sure to use the lightning arrester to 

comply EMC standards.

Note: When option code /C0 or /FB is specified, the 

conformity to the safety and EMC standards is 

excluded.

Note: When contactless AC switch is specified for the 

output-1 signal code or output-2 signal code, it will not 

comply with safety and EMC standards.

10.  ENVIRONMENT STANDARD

EU RoHS directive: EN IEC 63000

  (However, when contactless AC switch is specified for the 

output signal, and suffix code /C0 or /FB is specified, CE 

marking is not applicable because the product does not 

conform to safety and EMC standards.)

11.  TRANSPORT AND STORAGE CONDITIONS

 •

Temperature: −25 to 70°C

 •

Temperature change rate: 20°C per hour or less

 •

Humidity: 5 to 95%RH (no condensation)

 CAUTION

Keep this product in a conductive bag when plugged 

out, during transport or storage.

6

IM 77J01Q17-01E

6th Edition