Pulse output connection example

+

Open collector

Contactless AC switch 

(AC power supply connection)

Contactless AC switch 

(DC power supply connection)

9

7

Load

Load

Load

Load

+

9

7

Varistor or 

CR spark quenching circuit

Spark quenching diode

Output-2 is 

AC power

DC power

+

9

7

+

2 (+) 5 (

)

5 V, 12 V,  24 V

Voltage pulse

9

7

 CAUTION

 

Do not use output-2 for the isolated single-output type.

 

The power line and input/output signal lines should 

be installed away from noise-generating sources.  

Other wise accuracy cannot be guaranteed.

 

Make sure to earth ground the ground terminal 

through minimum resistance. The length and 

thickness of the grounding cable should be as 

short and thick as possible.  Directly connect the 

lead from the ground terminal (terminal no. 8) of 

the product to the ground.  Do not carry out daisy-

chained inter-ground terminal wiring.

 

The product is sensitive to static electricity; 

exercise care in operating it.  Before you operate 

the product, touch a nearby metal part to discharge 

static electricity.

 

The use of inductance (L) loads such as auxiliary 

relays and solenoid valves causes malfunction or 

relay failure; always insert a CR filter or diode for 

spark-removal into the line in parallel with the load. 

Recommended CR 

 • C: 0.5 to 1µF against contact current 1A

 •

R: 0.5 to 1Ω against contact voltage 1V

 

If the ambient temperature is 50 °C or more, please 

use the cable that the rated temperature is 70 °C or 

more.

4.  DESCRIPTION OF FRONT PANEL AND 

CONNECTION OF HANDY TERMINAL 

4.1  Front Panel 

The communications connector in the front panel is used for 

setting up parameters through the PC (VJ77 Parameters Setting 

Tool). The alarm-1 and alarm-2 LEDs light up if an alarm occurs 

(those LEDs are provided only when the output-2 is specified for 

alarm output (the output-2 suffix code is T).)

Alarm indicator lamp

(for alarm 1)

Communication connector

Alarm indicator lamp

(for alarm 2)

4.2  Connector for Communication

< How to connect with the setting tool>

JUXTA

VJ series

Modular-jack

conversion adapter

Dedicated adapter for VJ77 

(L4506HA)

Micro USB (USB2.0)

cable

PC with the

VJ77 installed

 • Use the VJ77 of version R2.02.01 or later.

3

IM 77J01Q17-01E

6th Edition