3.3. Mounting the FU20-FTS/MTS in
PR10 retractable
1 Take the sensor out of the box and apply
Teflon tape to the appropriate threaded
end.
2 Bind the separate wires of the cable
together with a piece of tape.
3 Take the fitting out of the box and remove
the option(s), if necessary.
4 Release the pigtail (cable gland)
completely. Do not undo the part in the
metal tube!
5 Lead the sensor cable through the tube
of the fitting, from the side where the
knurled knob has been removed. Attach
the sensor and cable as usual.
6 Hold the sensor still and turn the metal
tube onto the sensor. Don’t rotate the
cell, but rotate the tube of the fitting,
because the cable can be disconnected
from the cell, when rotating it.
7 Lead the loose part of the pigtail onto the
cable and screw it onto the fixed part.
8 Remove the tape.
FIG. 4 Dimensional drawing PR10...-D32 with mounted FU20 sensor