background image

IM 12B06J03-05E-E

17

Calibration of the FU20-FTS/MTS sensor 
has to be done with the pH analyzer 
connected. Normally the pH standards that 
are preprogrammed in the pH analyzer can 
be used for calibration with the pH analyzer 
set to 

“AUTOCAL”

. The FU20-FTS/MTS 

however is a differential pH sensor which 
needs pH buffers that have the same ionic 
strength because the sodium reference 
will change as the ionic strength changes. 
These pH buffers are not preprogrammed 
in the pH analyzer, which means the 
calibration has to be done with specific 
buffer solutions (see Section 6.1) and the 
pH analyzer set to 

“MANUAL”

 calibration. 

Refer to the analyzer Instruction Manual 
for details. Calibration data of the sensor is 
stored into the pH analyzer.

6.1 Buffer calibration

To calibrate the FU20-FTS/MTS sensor, 
two buffer solutions with known pH values 
are required. It is recommended that one 
buffer solution has a value near to pH 
7.00. Depending on the process value to 
be measured, the second buffer solution 
should be either acidic  
(below pH 7.00) or alkaline (above pH 7.00). 
Buffers which are available are: pH 4.00,  
pH 7.00 and pH 9.00. 
See Section 8 for ordering information.

The following is a very general 2-point manual 
calibration procedure with buffer solution:
1. Clean the sensor using a 5% solution of 

HCl;

2. Rinse sensor thoroughly with tap water  

(

DO NOT

 use demineralized water);

3. Immerse the sensor in the first buffer 

solution (pH 7.00 is recommended); 
Set the pH analyzer to 

“MANUAL” 

calibration. Wait until the pH reading is 
stable. Adjust the pH reading in the pH 
analyzer to the value indicated on the 
bottle (in this case 7.00). Go to solution 2 
to do a 2-point calibration;

4. Rinse sensor thoroughly with tap water;
5. Immerse the sensor in the second buffer  

(pH 4.00 is recommended); 
Wait until the pH reading is stable. Adjust 
the pH reading in the pH analyzer to the 
value indicated on the bottle (in this case 
4.00).  
Quit the manual calibration routine by 
accepting the new calculated calibration 
data of zero and slope.

After calibration, re-install the sensor into 
the process.

Note:

  It is important to understand it 

is possible the FU20-FTS/MTS 
differential sensor does not show 
the correct pH value in process after 
buffer calibration. The reason is that 
the ionic strength of the buffers is 1 
mol NaCl, which can deviate from the 
ionic strength of the process. For an 
accurate pH reading an extra 1 point 
process calibration has to be done 
(see Section 6.2) to compensate for 
differences in ionic strength. 

6.2 Process calibration

The following is a specific 1-point manual 
calibration procedure for the process to 
measure:
Set the pH analyzer to

 “MANUAL” 

calibration. Wait until the pH reading is 
stable. 
Adjust the pH reading in the pH analyzer 
to the value of the process sample that is 
analyzed using a laboratory measurement. 
For the laboratory measurement, we advice 
to use the Yokogawa PH72 personal pH 
meter. 
Quit the manual calibration routine by 
accepting the new calculated calibration 
data of zero. Slope value will be unchanged.

6.  GENERAL CALIBRATION & MAINTENANCE PROCEDURE

Summary of Contents for FU20-MTS

Page 1: ...IM 12B06J03 05E E 4th edition Instruction Manual Model FU20 FTS MTS Differential pH ORP sensor...

Page 2: ...IM 12B06J03 05E E 2...

Page 3: ...IM 12B06J03 05E E 3...

Page 4: ...IM 12B06J03 05E E 4...

Page 5: ...ON OF FU20 10 3 1 Typical installation 10 3 1 1 Preparing the sensor for use 10 3 1 2 Mounting the sensor 10 3 2 Mounting FU20 FTS using quick removal adapters 11 3 3 Mounting the FU20 FTS in PR10 ret...

Page 6: ...ode and Serial Number on the sensor are the same as on the packing list Also check any option s that were ordered are included and correct 1 3 Warranty and Service Yokogawa products and parts are guar...

Page 7: ...or year month numbering Table 1 Production Year code Table 2 Production Month code Year Year code Year Year code Month Month code 2014 P 2026 3 January 1 2015 R 2027 4 February 2 2016 S 2028 5 March 3...

Page 8: ...e element Pt1000 to IEC 751 Asymmetry potential 0 15 mV Linearity PH Slope 90 in pH 2 12 with pH pNa 2 NOTE The temperature sensor included in the FU20 FTS MTS is designed for cell compensation and fo...

Page 9: ...and control instruments ion selective electrodes Pressure Directive 97 23 EC as amended by Regulation EC no 1882 2003 Applying article 3 3 Sound Engineering Practice Damaging the screw thread of the s...

Page 10: ...plane to eliminate air bubbles in the pH glass bulb see Figure 1 Figure 1 Sensor installation 3 1 1 Preparing the sensor for use Remove the sensor from its shipping box and slide of the so called wet...

Page 11: ...2 Install the adapter in the process connection Tighten the adapter using a wrench on the adapter flats 3 Apply Teflon tape to the appropriate threaded end of the sensor 4 Place the O ring and screw t...

Page 12: ...t undo the part in the metal tube 5 Lead the sensor cable through the tube of the fitting from the side where the knurled knob has been removed Attach the sensor and cable as usual 6 Hold the sensor s...

Page 13: ...2B06J03 05E E 13 FU20 O ring Fitting Adapter FPS O ring HCNF SPRAY CLEANER FU20 FTS MTS ADAPTER FU20 FTS MTS FD40 Figure 5 Installation examples for the FU20 Process Flow Cleaning Solution water acid...

Page 14: ...0 2 5 0 0 2 25 0 1 Figure 7 Dimensions 1 FU20 FTS MTS adapter Stainless Steel Titanium and FU20 FTS MTS adapter for FF40 FS40 and FD40 fittings 4 DIMENSIONS Figure 6 Dimensions FU20 FTS MTS FU20 VP T...

Page 15: ...Yokogawa pH instrument Differential pH pNA ORP rH 15 16 13 17 11 12 14 63 3 none Differential ORP for pH 13 17 11 12 14 15 63 3 none Differential ORP 13 17 11 12 14 15 63 3 none pH compensated for pH...

Page 16: ...ransparent Coax 2 Transparent Coax 2 Clear 14 Yellow 15 Transparent Coax 1 Yellow Yellow 16 Brown 17White Transparent Coax 2 Transparent Coax 2 Brown Note For ORP measurement the EXA FLXA terminals 13...

Page 17: ...commended Set the pH analyzer to MANUAL calibration Wait until the pH reading is stable Adjust the pH reading in the pH analyzer to the value indicated on the bottle in this case 7 00 Go to solution 2...

Page 18: ...FU20 FTS MTS sensor does not have an open reference junction and is hermetically sealed from the process it does not suffer from poisoning diffusion and fouling This means the FU20 FTS MTS sensor requ...

Page 19: ...del NPT PPS body Tapered Thread Dome shaped FSM PPS body Tapered Thread Flat Surface FTD PVDF body Tapered Thread Dome shaped FTS PVDF body Tapered Thread Salt Sensitive membrane Silicone sealing MTS...

Page 20: ...F10 xxx F Extension cable with terminated ends Specify length in whole meters e g XXX 005 010 025 050 meter WU10 V D 02 Cable for Variopin dual coax 2 meter WU10 V D 05 Cable for Variopin dual coax 5...

Page 21: ...I chloride 50 O O O O O O O O O O O O O O O O O O Sodium sulfite sat O O O O O O O O O O O O O O O O O O Sodium carbonate sat O O O O O O O O O O O O O O O O O O O O O Potassium chloride sat O O O X X...

Page 22: ...IM 12B06J03 05E E 22...

Page 23: ...IM 12B06J03 05E E 23...

Page 24: ...etherlands www yokogawa com eu YOKOGAWA ELECTRIC CORPORATION World Headquarters 9 32 Nakacho 2 chome Musashino shi Tokyo 180 8750 Japan www yokogawa com YOKOGAWA CORPORATION OF AMERICA 2 Dart Road New...

Reviews: