<4. Setting converter>
4-6
IM 12A01G01-03EN
1st Edition : Mar. 25, 2021-00
4.1
mA input settings (Ad)
First select the application of mA input from “Current”, “Temperature”, “pH” or “Conductivity””,
then set 0% value, 100% value of each, according to the application.
This function is available when mA input/output is “-N4” in the suffix code. When mA input/output
is “-N2”, mA input is not settable.
If you do not set mA input (Ad), set the application to “None”.
Menu
Parameter
default
Range
lower limit*1
upper limit *2
mA input setting (Ad) Damping time
0.0 s
0 s
3600 s
AI upper limit
20 mA
3.8 mA
20.5 mA
AI lower limit
4 mA
3.8 mA
20.5 mA
Temperature 4 mA
0°C
-40.0°C
260.0°C
Temperature 20 mA 100°C
-40.0°C
260.0°C
pH 4 mA
0 pH
-2
16
pH 20 mA
14 pH
-2
16
Conductivity 4 mA
0 S/m
0 S/m
1 S/m
Conductivity 20 mA 0.05 S/m
0 S/m
1 S/m
*1:
Configurable lower limit
*2:
Configurable upper limit
4.1.1
Type
Select mA input application out of “None”, “Current”, “Temperature”, “pH”, “Conductivity”.
4.1.2
Damping time
This is the time taken for a response to a step input change to reach 90% of the final value
(attenuation time). The unit is second.
4.1.3
AI Upper Limit/AI Lower Limit
Set AI upper/lower limit for error detection of upper/lower limit. The lower limit must be lower than
the upper one.
4.1.4
Temperature (4 mA/20 mA)
Set each value for 0% (4 mA) and 100% (20 mA) to use mA input as temperature parameter. The
value has to be different from each other.
4.1.5
pH (4 mA/20 mA)
Set each value for 0% (4 mA) and 100% (20 mA) to use mA input as pH parameter. The value
has to be different from each other.
4.1.6
Conductivity (4 mA/20 mA)
Set each value for 0% (4 mA) and 100% (20 mA) to use mA input as conductivity parameter. The
value has to be different from each other.