<4. Setting converter>
4-21
IM 12A01G01-03EN
1st Edition : Mar. 25, 2021-00
Table 4.5 Error list
Location
Alarm
Alarm name
default*1
System
I/O Alarm
Fail safe occur
F
Maintenance status Outputs in HOLD 1/2/3/4
N
mA1/2/3/4 Output Maint. status
N
Setting alarm
mA configuration error 1/2/3/4
F
Display 1/2/3/4 setting error
C
Auto wash/cal. setting error
C
Auto start date error
C
Contact config. error 1/2/3/4
F
pH
Device status
Sensor communication error
F
Measure alarm
Temperature too high
S
Temperature too low
S
pH too high
S
pH too low
S
Imped. (pH/ORP) too high
M
Imped. (pH/ORP) too low
M
Impedance (ref) too high
M
Impedance (ref) too low
M
SSA temp outside spec
S
Sensor status
Sensor not detect
F
Temp element not detect
F
ID chip failure
F
LE not detect
F
Calibration Due
N
imp1 glassbreak detect
F
imp2 glassbreak detect
F
SC
Device status
Sensor communication error
F
Measure alarm
Temperature too high
S
Temperature too low
S
Conductivity too high
S
Conductivity too low
S
Temp comp1 warning
S
Polarization detected
N
SSA temp outside spec
S
Sensor status
Sensor not detect
F
Temp element not detect
F
ID chip failure
F
Calibration Due
N
TB
Device status
Sensor communication error
F
Measure alarm
Tubidity high
S
Turbidity low
S
Photodetector current error
S
Light source failure
S
Auto Zero cal. result error
S
Auto cal. Unstablet
S
Sensor status
Humidity sensor failure
S
CL
Device status
Sensor communication error
F