Tech
Note:
TNA1403
Date:
February
5,
2014
FLXA21-P1 Analyzers
Programming Custom Buffer Tables
FLXA21, comes with NIST Buffer tables pre-programed intot the analyzer. When using the
autocalibration mode, the system uses these preprogramed tables for reference.
However there is an option for customers to change buffer table data information.
There are three sets of NIST buffer tables, pH 4.01, pH 6.98, and pH9.18, by using the free
program options, you can choose to change whichever table and however many
tables you need to change. This document is intended to assist customers with the
steps that need to be taken inorder to to change the Buffer Tables.
1.)
Click on the settings icon (wrench)
2.)
Using either the the scroll key or by
clicking directly on the diamond next to
Commissioning, select it.