YOKOGAWA FD20 Instruction Manual Download Page 12

IM 12B06K02-01E-E

12 

3. Installation

3.1 Selecting the installation site

It is important that, whatever method of 

mounting is used, the point of measurement is 

truly representative for the entire solution.

-  Avoid an area where the measurement value 

varies significantly.

-  Ensure that the sensors are always immersed 

in the process liquid (not dry).

-  When selecting the installation site, first check 

that the liquid temperature and pressure 

meets the sensors and fittings requirements.

-  If the fitting is used in a tank with stirrer  

or agitator, or if it is placed in a fast flowing 

process, care must be taken that it  

is adequately supported to prevent any 

movement.

-  Install the fitting in a site where it can be easily maintained. 

The mounting position of the sensors must be within the area as shown in figure 3.1.

3.2 Installation

Before installing the immersion fitting, check that all parts and accessories were supplied by 

Yokogawa.

Ensure that there is enough spare around to withdraw the fitting for calibration and 

maintenance.

3.2.1 “Hoisting eye” type mounting

On top of each fitting is a “hoisting eye” to withdraw from the process.

3.2.2 Flange mounting fitting

If ordered, the immersion fittings are supplied with a flange for fixing in a tank. The flange is 

connected to the fitting as shown in the dimensional drawings. These drawings also show 

the configuration of holes.

Table 2.  Flanges being used as a standard

INCORRECT

15º

15º

HORIZONTAL

GOOD

GOOD

Fig. 3.1 Mounting position of sensors

Type  

Material  TypePP 

Material 

DN 

Thickness 

fitting fitting flange 

flange 

 

flange

1-hole  

PVC 

glue-flange 

PVC 

32

3-hole  

SS 

weld-flange 

SS 

80 

20 mm

4-hole  

SS 

weld-flange 

SS 

100 

20 mm

3-hole  

PP 

collar-bush + 

PP + 

80 

 

   

 

lap-joint flange 

PP (steal inlay)

4-hole  

PP 

collar-bush + 

PP + 

125 

 

   

 

lap-joint flange 

PP (steal inlay)

3-hole  

PVDF 

collar-bush + 

PVDF + 

80 

 

   

 

lap-joint flange 

PP (steal inlay)

4-hole  

PVDF 

collar-bush + 

PVDF + 

125 

   

 

lap-joint flange 

PP (steal inlay)

IM12B06K02-01E-E_ed09.indd   12

07/03/16   09:51

Summary of Contents for FD20

Page 1: ...IM 12B06K02 01E E 9th edition Instruction Manual Model FD20 Immersion Fittings pH ORP YOKOGAWA IM12B06K02 01E E_ed09 indd 1 07 03 16 09 51...

Page 2: ...IM 12B06K02 01E E 2 IM12B06K02 01E E_ed09 indd 2 07 03 16 09 51...

Page 3: ...pe mounting 12 3 2 2 Flange mounting fitting 12 3 3 Cables and sensor mounting 13 3 3 1 1 Hole fitting 13 3 3 2 3 Hole and 4 hole fitting 15 4 Mounting of Accessoires 16 4 1 Mounting kit option B FP20...

Page 4: ...IM 12B06K02 01E E 4 IM12B06K02 01E E_ed09 indd 4 07 03 16 09 51...

Page 5: ...shipment from the manufacturer Individual Sales organizations can deviate from the typical warranty period and the conditions of sale relating to the original purchase order should be consulted Damage...

Page 6: ...tion Year code Year Year code Year Year code 2014 P 2026 3 2015 R 2027 4 2016 S 2028 5 2017 T 2029 6 2018 U 2030 7 2019 V 2031 8 2020 W 2032 9 2021 X 2033 A 2022 Y 2034 B 2023 Z 2035 C 2024 1 2036 D 2...

Page 7: ...s are not included The noted weights are at an immersion length of 1 m Mounting by means of the hoisting eye or flange mounting 2 2 Functional specifications Temperature min 10 C max depending on mate...

Page 8: ...1 electrode mounting holes Immersion Between 5 and 20 in dm length 2 example 06 0 6 m Flange 3 NF No flange Working F1 Flange DN32 for 1 hole fitting pressure F2 Flange DN80 for 3 hole fitting PP not...

Page 9: ...12B06K02 01E E 9 2 4 External dimensions 90 75 ISO 228 2 G 1 2 2x k D 125 b dg dg D k b 30 25 40 L ISO 228 2 G1 2 1 hole PVC 3 hole PP and PVDF Flange Flange IM12B06K02 01E E_ed09 indd 9 07 03 16 09 5...

Page 10: ...k D b 35 dg L 35 L 130 110 M8 ISO 228 2 G1 2 3x L 76 80 ISO 228 2 G 1 2 2x 80 10 3 hole SS b 188 k D L dg 4 hole PP and PVDF IM 12B06K02 01E E Flange Flange IM12B06K02 01E E_ed09 indd 10 07 03 16 09 5...

Page 11: ...20dm 200 160 18 20 FD20 P37 S2 5 20dm 190 152 4 20 18 FD20 F37 S3 5 20dm 190 152 4 20 18 FD20 S37 NF 5 20dm FD20 S37 F4 5 20dm 200 160 18 20 FD20 S37 S4 5 20dm 228 6 190 5 19 23 8 FD20 P47 NF 5 20dm F...

Page 12: ...ts and accessories were supplied by Yokogawa Ensure that there is enough spare around to withdraw the fitting for calibration and maintenance 3 2 1 Hoisting eye type mounting On top of each fitting is...

Page 13: ...s shown in figure 3 3 Fig 3 2 Electrode cables Model code Suffix code Description WU20 D Electrode cable Type PC COAX LT TRIAX Length in mm 01 1 mtr 02 2 mtr 05 51 2 mtr 10 10 mtr 15 15 mtr 20 20 mtr...

Page 14: ...hole fitting O rings 11x3 Electrode cable model WU20 Attention Ensure that the cable has slack remaining in the fitting so that sensor replacement is simplified To prevent damage of the O rings the nu...

Page 15: ...w end see figure 3 4 Prepare the sensors to be used in accordance with the directions for use and mount them in the fitting as shown in fig 3 4 For stable measurements the glass and reference electrod...

Page 16: ...of the electrodes see seperate sheets Parts and accessories Order nr Description FP20 S13 Mounting kit for SR20 AC32 10x K1500HC Rubber rings for FP20 S13 10x K1500GE O rings for SR20 D AC32 electrode...

Page 17: ...ows the mounting Note The pressure on the KCI solution must always exceed the liquid pressure The refillable electrodes can be connected to a KCI reservoir via a silicone tubing see figure 4 2 The pre...

Page 18: ...of cleaning systems see figure 4 3 brush cleaning electrically or pneumatically driven chemical cleaning 4 3 1 Selection criteria Cleaning system Mechanical Chemical Applications with Brush Acid Base...

Page 19: ...tem wipes across the glass membrane of the electrode every 30 seconds Electrically or pneumatically driven versions are available Figures 4 4a and 4 4b show the mounting Control unit Parts and accesso...

Page 20: ...n the liquid to be measured The cleaning period and frequency can be adjusted on a control unit The mounting procedure is shown in figure 4 5 Parts and accessories Order nr Description K1520FJ Tubing...

Page 21: ...ng set PPS Ryton O ring silicone Connection nylon Weight approx 300 g Mounting wall mounting support with hole for screw M5 Temperature pressure ratio max 200 kPa 2 bar at 100 C Parts and accessories...

Page 22: ...S K1500FC EPDM Set O rings for 3 hole fitting SS K1500GX Set O rings for 4 hole fitting SS K1500FH viton Set O rings for 4 hole fitting SS K1500FG EPDM Set O rings for 4 hole fitting SS K1500DQ 10 m p...

Page 23: ...IM12B06K02 01E E 23 1 Hole PVC Note This fitting has no liquid earth connection 3 Hole PP and PVDF 4 Hole PP and PVDF 3 Hole SS 4 Hole SS IM12B06K02 01E E_ed09 indd 23 07 03 16 09 51...

Page 24: ...awa com eu YOKOGAWA ELECTRIC CORPORATION World Headquarters 9 32 Nakacho 2 chome Musashino shi Tokyo 180 8750 Japan www yokogawa com YOKOGAWA CORPORATION OF AMERICA 2 Dart Road Newnan GA 30265 USA www...

Reviews: