<9. General Specifications>
9-3
IM 01C25K01-01E
EMC Conformity Standards
EN 61326-1 Class A, Table2 (For use in
industrial locations)
EN 61326-2-3
EN 61326-2-5 (for Fieldbus)
Immunity influence during the test
Differential pressure: Output shift is specified
within ±1% of 1/20 Max span.
Static pressure: Output shift is specified within
±2% of 1MPa span.
Status Output Line*: Continues to operate
without reversal.
*: Only for /AL option.
Condition for EMC test for F
OUNDATION
fieldbus
type: The shield of the cable and the case
are connected with a capacitor of 10nF.
Physical Specifications
Wetted Parts Materials
Diaphragm, Cover Flange, Process Connector,
Capsule Gasket, Vent/Drain Plug, Manifold,
Orifice, Spacer, and Orifice gasket
Non-wetted Parts Materials
Bolting
ASTM-B7 carbon steel, 316L SST stainless
steel, or ASTM grade 660 stainless steel
Housing
Low copper cast aluminum alloy with
polyurethane paint or ASTM CF-8M stainless
steel
Degrees of Protection
IP66/IP67, Type 4X
Cover O-rings
Buna-N, fluoro-rubber (option)
Nameplate and tag
316SST (including /N4 wired tag)
Fill Fluid
Silicone, Fluorinated oil (option)
Weight
[Installation code 7, 8, and 9 and measurement
span code M and H]
4.5 kg (9.9 lb) without integral indicator and
mounting bracket.
[Installation code 7, 8 and 9 and measurement
span code F]
5.4 kg (11.9 lb) without integral indicator and
mounting bracket.
Add 1.5 kg (3.3lb) for amplifier housing code 2.
Connections
Refer to “Model and Suffix Code.”
Process Connection of Cover Flange:
IEC61518