2-8
IM 701610-61E
2.4 Analyzing/Searching Data
• Setting the Reference Point
When Setting the Trigger Position to the Analysis Reference Point
29.
Turn the jog shuttle to select Trigger Position.
30.
Press SELECT. The button to the left of Trigger Position is highlighted and the
analysis reference point is set to the trigger position. Proceed to step 33.
When Manually Setting the Analysis Reference Point
29.
Turn the jog shuttle to select Manual.
30.
Press SELECT. The button to the left of Manual is highlighted.
31.
Turn the jog shuttle to select the box to the right Manual.
32.
Set the value in a similar fashion as described in steps 8 through 10. Proceed
to step 33.
Note
If you set the display to translucent mode when manually setting the analysis reference point,
you can set the analysis reference point while viewing the waveform display. For details on
the translucent mode display, see section 8.7, “Turning Translucent Mode ON/OFF” in the
DL1620/DL1640/DL1640L User’s Manual IM 701610-01E
.
• Setting the Bit Order
33.
Turn the jog shuttle to select Bit Order.
34.
Press SELECT to select MSB First or LSB First.
Press ESC to close the dialog box.
• Executing the Analysis
35.
Press the Analyze Exec soft key. The waveform is analyzed, and the result is
displayed on the right side of the screen.
If you select the analysis number by pressing the List soft key and turning the
jog shuttle, the data corresponding to the selected analysis number is displayed
expanded on the ZOOM display.
Display this data point
on the ZOOM display
Analysis result
ZOOM display
Note
If the analysis result contains indefinite data, “*” is displayed in the Data1 and Data2 boxes.