1-21
IM 701610-61E
1
I
2
C-Bus Signal Analysis Function
Setting the Search Conditions
32.
Press the Data Search soft key.
33.
Press the Search Setup soft key. The Search Setup dialog box appears.
34.
Turn the jog shuttle to move the cursor to Type.
35.
Press SELECT to select Byte Pattern or Indefinite State.
• When Byte Pattern (Pattern Search) Is Selected
Setting the Address Pattern
36.
Turn the jog shuttle to select Address Pattern.
37.
Press SELECT. The button to the left of Address Pattern is highlighted.
38.
Turn the jog shuttle to move the cursor to the box you wish to set the Address
Pattern.
39.
Press SELECT to display the entry box.
40.
Turn the jog shuttle to set the value. Bits set to “X” will not be used as a search
condition. The search conditions can be specified using binary (BIN) or
hexadecimal (HEX) values. Press SELECT or ESC to close the entry box. The
value will be confirmed.
41.
Turn the jog shuttle to select Ack.
42.
Press SELECT to set the Ack to “X,” “1,” or “0”.
Setting the Data Pattern
43.
Turn the jog shuttle to select Data Pattern.
44.
Set the data pattern in a similar fashion as described in steps 37 through 40.
Press ESC to close the dialog box. Proceed to step 45.
Note
•
When using binary (BIN) display, set the LSB of the address pattern using “R” (Read), “W”
(Write), and “X.”
•
When setting the pattern by using binary (BIN) display, the entry box is not displayed.
• When Indefinite State (Indefinite Data Search) Is Selected
You do not have to set any items.
Press ESC to close the dialog box. Proceed to step 45.
1.5 Analyzing/Searching Data