45

Voltage Source/Measurement Specifications

Range

Measurement and Source Range

Resolution

Accuracy (1 year)

Notes

of Setting

of Reading

Offset

90 mV

–11.000 to +99.999 mV

1 μV

0.015%

10 μV

Maximum output current: 1 mA 

At the ambient temperature (Ta) of 23±5°C.
For the accuracies of other  ambient temperature ranges (Ta<18°C, Ta>28°C), add the temperature coefficient 0.005% of Range/°C.

Common Measurement Section Specifications

Maximum input across terminals: 42 Vpeak
Maximum input to ground: 42 Vpeak

Common Source Section Specifications

Output resistance: 40 mΩ or less
Output response: 300 ms or less
Maximum load: C≤0.1 μF, L≤10 mH
Maximum input across terminals: 42 Vpeak
Maximum input to ground: 42 Vpeak

Thermocouple Accuracy Equation

Accuracy with respect to the measurement or source temperature (t) is expressed as constant or a linear equation of t.
Example: Accuracy for Thermocouple K (terminal A) measurement value of 1000.0°C = ±(0.5 + (1000.0-500)×0.02%)°C = ±0.6°C

Specifications

Summary of Contents for CA310

Page 1: ...User s Manual IM CA310 02EN 3rd Edition CA310 Volt mA Calibrator CA320 TC Calibrator CA330 RTD Calibrator Getting Started Guide...

Page 2: ...as manuals for the CA310 CA320 CA330 Please read all manuals Manual Title Manual No Description CA310 Volt mA Calibrator CA320 TC Calibrator CA330 RTD Calibrator Getting Started Guide IM CA310 02EN T...

Page 3: ...ons or find any errors please contact your nearest YOKOGAWA dealer Copying or reproducing all or any part of the contents of this manual without the permission of YOKOGAWA is strictly prohibited Trade...

Page 4: ...you for purchasing YOKOGAWA products YOKOGAWA provides registered users with a variety of information and services Please allow us to serve you best by completing the product registration form accessi...

Page 5: ...ealer CA310 CA320 CA330 Check that the product that you received is what you ordered by referring to the model name on the name plate on the rear panel of the main unit For reference the model name an...

Page 6: ...Binding post red red pair 99046 1 One short plate included Batteries 4 4 4 AA alkaline batteries LR6 Portable case B9108NK 1 1 1 Antiskid rubber 1 1 1 Antiskid rubber for the bottom panel Manuals IM C...

Page 7: ...AC Operating temperature range 10 C to 40 C RJ sensor 90080 P PT100 JIS AA class or equivalent Thermometer operating temperature range 10 C to 55 C 8 polar miniDin connector Shielded cable 1 5 m Y ter...

Page 8: ...voltage current measuring instrument CA310 thermocouple measuring instrument CA320 and RTD measuring instrument CA330 Use this instrument only as a calibrator Check the Physical Appearance Do not use...

Page 9: ...outdoors or in locations subject to rain or water Or use the instrument in such locations Install the instrument so that you can immediately turn off the power if an abnormal or dangerous condition oc...

Page 10: ...l apparence physique Ne pas utiliser l instrument si son int grit physique semble tre compromise Adaptateur c a Utiliser exclusivement l adaptateur c a d di pour l instrument Batteries Ne pas m lange...

Page 11: ...t vous vous exposeriez des risques de choc lectrique ou de br lure ou vous risqueriez d endommager l quipement Mettez toujours le dispositif mesurer hors tension avant de le connecter l quipement Il e...

Page 12: ...nd precautions that can be taken to prevent such occurrences French AVERTISSEMENT Attire l attention sur des gestes ou des conditions susceptibles de provoquer des blessures graves voire mortelles et...

Page 13: ...stic household waste EU Battery Directive This directive is valid only in the EU Batteries are included in this product This marking indicates they shall be sorted out and collected as ordained in the...

Page 14: ...y 17 Making Preparations for Measurements 18 Handling Precautions 18 Installing the Instrument 20 Connecting Cables 23 Using Thermocouple Mini Plugs CA320 only 25 Installing and Removing Batteries 26...

Page 15: ...ry switch Selects the range and sensor type You can also select the setup mode SETTING Measurement source switch POWER key Turns the power on and off AC adapter jack Connect the AC adapter an accessor...

Page 16: ...ltage corresponding to temperature CA320 or resistance CA330 span display RJ sensor temperature CA320 Pauses or resumes sweeping in sweep mode Set the output to 0 in span check Press simultaneously du...

Page 17: ...f if LOOP POWER is in use TYPE key Select the sensor CA320 CA330 Press simultaneously to switch the sub display item voltage corresponding to temperature span display RJ sensor temperature CA320 On th...

Page 18: ...Step sweep Linear sweep 9 Main display unit 10 Calibration mode not displayed normally 11 Appears when auto power off is enabled 12 Main display 13 Selection status of the TC connection terminal When...

Page 19: ...s are subjected to mechanical shock they may be damaged The instrument may not perform measurements correctly due to damage or deformation that is not visible to the naked eye Do Not Scratch the LCD B...

Page 20: ...or a long time Storage Precautions Avoid the following kinds of places for storing the instrument Where the temperature falls outside the storage temperature and humidity ranges In direct sunlight or...

Page 21: ...the instrument when it is wet Doing so may damage the instrument French AVERTISSEMENT Installer l instrument de mani re pourvoir imm diatement le d brancher du secteur en cas de fonctionnement anorma...

Page 22: ...Condensation may occur if the instrument is moved to another place where the ambient temperature or humidity is higher or if the temperature changes rapidly In such cases before you use the instrumen...

Page 23: ...V For measurements performed on power source circuits Entrance cables cable systems etc Portable Case and Antiskid Rubber How to Use the Portable Case 1 Unfasten the hook on the strap end and hooks on...

Page 24: ...smitter 500 mV 5 V 30 V or 50 V range 20 mA or 50 mA range LOOP POWER DC voltage source DC current source Hart Communicator BRAIN Terminal CA310 source 20 mA range mA SIMULATE Distributor A V DC volta...

Page 25: ...from the binding post The Lo banana terminal and the Lo thermocouple mini plug terminal are shorted internally CA330 measurement 4 W CA330 source SEMSE Lo INPUT Lo RTD INPUT Hi SENSE Hi 3W Lo Hi 4W Sh...

Page 26: ...em to be calibrated thermocouple or range of the device to be calibrated Use the thermocouple mini plug set 90040 or 90045 or a thermocouple mini plug that you prepared To connect the plug and the ite...

Page 27: ...jamais remplacer les batteries lorsque la mesure est en cours Inserer les batteries en observant la polarite correcte A defaut les batteries risquent de fuir de chauffer ou d eclater Ne pas m langer...

Page 28: ...adaptateur c a lorsque le bo tier principal est en marche Utiliser uniquement l adaptateur de CA d di et le c ble d alimentation de cet instrument Ne posez pas d objets sur l adaptateur c a ni sur le...

Page 29: ...tion When the power is turned on a memory check starts automatically If the instrument does not power on properly turn the power off and check that The dry cells are inserted properly New and old dry...

Page 30: ...and SOURCE Source Mode Use the measure source switch to select MEASURE measurement mode or SOURCE source mode Note On the CA320 or CA330 if you select SOURCE the display shows ON and output starts On...

Page 31: ...of span Set to 100 of span Set to 0 of span Press simultaneously to set the output mode On the CA310 press the OUTPUT ON OFF key to start or stop sweeping On the CA320 CA330 press the TYPE key to do...

Page 32: ...IN communication resistance Temp measurement source terminal RJ sensor banana terminal Burnout detection Temperature reference Extended sensor Initializes the span setting Initializes the settings Opt...

Page 33: ...ck that the remaining battery power is sufficient 17 The screen is dark Turn the backlight on 31 The screen turns off Set auto power off to OFF 31 The display is odd Confirm that the ambient temperatu...

Page 34: ...Set to a value less than the span 100 value If the span 100 value is set less than or equal to the span 0 value an error occurs Set to a value greater than the span 0 value 15 Recommended Replacement...

Page 35: ...to 60 000 mA 1 A 0 015 3 A Input resistance 10 or lower LOOP POWER 20 mA 0 000 to 24 000 mA 1 A 0 015 3 A Supply voltage 24 V 1 V when HART communication resistance is off Maximum load current 24 mA 2...

Page 36: ...015 3 A Compliance voltage 24 V 20 mA SIMULATE 0 000 to 24 000 mA 1 A 0 015 3 A External power supply 5 V to 28 V At the ambient temperature Ta of 23 C 5 C For the accuracies of other ambient tempera...

Page 37: ...0 x 2 0 IEC60584 1 200 0 C t 0 0 C 0 5 t x 0 3 0 5 t x 0 3 0 0 C t 500 0 C 0 5 0 5 500 0 C t 1000 0 C 0 5 t 500 x 0 02 0 5 t 500 x 0 02 J 210 0 C t 0 0 C 0 5 t x 0 3 0 5 t x 0 3 IEC60584 1 0 0 C t 120...

Page 38: ...000 x 0 06 1 0 t 1000 x 0 06 Extra TC D W3Re W25Re 0 0 C t 300 0 C 1 4 1 8 ASTM E1751 E1751M 09e1 300 0 C t 1500 0 C 1 2 1 2 1500 0 C t 2315 0 C 1 8 2 2 G W W26Re 100 0 C t 300 0 C 1 4 1 8 ASTM E1751...

Page 39: ...0 1 0 J 210 0 C t 0 0 C 1 0 t x 0 5 1 0 t x 0 5 IEC60584 1 0 0 C t 1200 0 C 1 0 t x 0 02 1 0 t x 0 02 T 250 0 C t 200 0 C 2 5 t 200 x 7 0 2 5 t 200 x 7 0 IEC60584 1 200 0 C t 0 0 C 1 t x 0 75 1 t x 0...

Page 40: ...Re 0 0 C t 300 0 C 1 4 1 8 ASTM E1751 E1751M 09e1 300 0 C t 1500 0 C 1 2 1 2 1500 0 C t 2315 0 C 1 8 2 2 G W W26Re 100 0 C t 300 0 C 1 4 1 8 ASTM E1751 E1751M 09e1 300 0 C t 1500 0 C 1 2 1 2 1500 0 C...

Page 41: ...02 0 7 t 500 x 0 02 J 210 0 C t 0 0 C 0 7 t x 0 3 0 7 t x 0 3 IEC60584 1 0 0 C t 1200 0 C 0 7 t x 0 02 0 7 t x 0 02 T 250 0 C t 200 0 C 1 7 t 200 x 5 0 1 7 t 200 x 5 0 IEC60584 1 200 0 C t 0 0 C 0 7 t...

Page 42: ...00 0 C 1 2 1 2 1500 0 C t 2315 0 C 1 8 2 2 G W W26Re 100 0 C t 300 0 C 1 4 1 8 ASTM E1751 E1751M 09e1 300 0 C t 1500 0 C 1 2 1 2 1500 0 C t 2315 0 C 1 8 2 2 PLATINEL II 0 0 C t 100 0 C 0 6 1 8 ASTM E1...

Page 43: ...r 90080 sold separately Specifications Sensor used RTD Pt100 four wire system Standalone accuracy JIS AA class or equivalent Specified current 1 mA Operating temperature range 10 C to 55 C Y terminal...

Page 44: ...0 0 C t 1000 0 C 0 3 t 500 x 0 02 0 3 t 500 x 0 02 J 210 0 C t 0 0 C 0 3 t x 0 2 0 3 t x 0 2 IEC60584 1 0 0 C t 1200 0 C 0 3 t x 0 02 0 3 t x 0 02 T 250 0 C t 200 0 C 0 7 t 200 x 1 0 0 7 t 200 x 1 0 I...

Page 45: ...2500 0 C 0 8 t 1000 x0 06 0 8 t 1000 x0 06 Extra TC D W3Re W25Re 0 0 C t 300 0 C 1 2 1 6 ASTM E1751 E1751M 09e1 300 0 C t 1500 0 C 1 0 1 0 1500 0 C t 2315 0 C 1 6 2 0 G W W26Re 100 0 C t 300 0 C 1 2 1...

Page 46: ...Common Measurement Section Specifications Maximum input across terminals 42 Vpeak Maximum input to ground 42 Vpeak Common Source Section Specifications Output resistance 40 m or less Output response...

Page 47: ...o 3 mA JIS C 1604 1989 JPt100 0 0 C t 510 0 C 0 3 t x 0 033 0 3 t x 0 033 3926 200 0 C t 0 0 C 0 3 0 3 0 1 to 3 mA Minco Application Aid 18 0 0 C t 630 0 C 0 3 t x 0 033 0 3 t x 0 033 PT200 3851 200 0...

Page 48: ...the accuracies of other ambient temperatures Ta 18 C Ta 28 C add the temperature coefficient 0 05 C C Source measured value display resolution 0 1 C The above figures are accuracies through four wire...

Page 49: ...of all cables are the same Assume the accuracies through two wire system measurement to be the same as those through three wire system measurement However cable resistance are not taken into consider...

Page 50: ...or lower CA320 Approx 55 hours CA330 Approx 55 hours For all cases backlight off Auto power off Approx 20 minutes disabled when the separately sold AC adapter is connected Measurement display update i...

Page 51: ...nity test environment within 10 of the range Environmental standard Compliant standard EN50581 Monitoring and instruments 1 The overvoltage category installation category is a value used to define the...

Page 52: ...IM CA310 02EN 51 External Dimensions 92 42 192 CA310 CA320 CA330 Unless otherwise specified tolerances are 3 however tolerances are 0 3 mm when below 10 mm Unit mm Specifications...

Reviews: