30
IM CA310-02EN
Setting the Output Mode
There are four output modes: normal output, span check, step sweep, and linear sweep. Press the keys simultaneously in the
following figure to select normal output (no display),
(span check), (step sweep), (linear sweep).
▲
▲
▲
▲
▲
▲
▲
▲
Pauses or resumes sweeping
Set the output value of span check.
▲
▲
▲
▲
▲
▲
▲
▲
±1 count
±100 counts
±25% steps of span
Set to 100% of span
Set to 0% of span
Press simultaneously to
set the output mode.
On the CA310, press the
OUTPUT ON/OFF
key to start or stop sweeping. On the CA320/CA330, press the
TYPE
key to do the same.
When sweeping is stopped or paused, you can press the left-most key to set the output span to 0%.
CAUTION
Be careful that the output value does not exceed the allowable input value of the target device.
French
ATTENTION
Veillez à éviter que la valeur de sortie ne dépasse la tolérance d’entrée de la cible de sortie.