4-7
IM AQ7277-17EN
Commands
:ACQuire:AVERage:STOP
Function Stops the averaging measurement.
Syntax
:ACQuire:AVERage:STOP
Example
:ACQuire:AVERage:STOP
:ACQuire:AVERage:TIME
Function Sets the average time or queries the current
setting.
Syntax
:ACQuire:AVERage:TIME {<NRf>|AUTO}
:ACQuire:AVERage:TIME?
<NRf> = 2 to 1800
Example
:ACQuire:AVERage:TIME AUTO
:ACQuire:AVERage:TIME 1200
:ACQuire:AVERage:TIME? ->
:ACQ:AVER:TIME 60
Description The following value can be set.
Auto: AUTO
2 sec:
2
5 sec:
5
10 sec:
10
20 sec:
20
30 sec:
30
1 min:
60
3 min:
180
5 min:
300
10 min:
600
20 min:
1200
30 min:
1800
:ACQuire:AVERage:TYPE
Function Sets the average unit or queries the current
setting.
Syntax
:ACQuire:AVERage
:TYPE {TIMES|DURATION}
:ACQuire:AVERage:TYPE?
Example
:ACQuire:AVERage:TYPE TIMES
:ACQuire:AVERage:TYPE DURATION
:ACQuire:AVERage:TYPE? ->
:ACQ:AVER:TYPE DURATION
Description Times:
TIMES
Duration: DURATION
:ACQuire:DRANge
Function Sets the distance range or queries the current
setting.
Syntax
:ACQuire:DRANge {<NRf>|AUTO}
:ACQuire:DRANge?
<NRf> = 500m to 400000m
(0.5km to 400km)
Example
:ACQuire:DRANge AUTO
:ACQuire:DRANge 500
:ACQuire:DRANge 500m
:ACQuire:DRANge 5000
:ACQuire:DRANge 100km
:ACQuire:DRANge? -> :ACQ:DRAN 1000
Description The value that you can specify varies depending
on the wavelength setting.
The selectable distance range varies depending
on the wavelength
Setting the distance range
Auto: AUTO
500 m:
500
1 km:
1000
2 km:
2000
5 km:
5000
10 km:
10000
20 km:
20000
50 km:
50000
100 km:
100000
200 km:
200000
300 km:
300000
400 km:
400000
:ACQuire:OFFSet
Function Sets the horizontal measurement start position or
queries the current setting.
Syntax
:ACQuire:OFFSet {<NRf>}
:ACQuire:OFFSet?
Example
:ACQuire:OFFSet 1000
:ACQuire:OFFSet? ->
v:ACQ:OFFS 999.00
Description The unit is set to m. (meter)
4.3 Device-Specific Commands
1
2
3
4
5
App
Index