3-8
IM 704420-01E
3.5
Connecting the Cable or Probe
Position of the Signal Input Connector
The signal input connector is located at the lower right section of the front panel.
Connect a cable or a probe with a BNC connector.
CLOCK
RF
BI-PHASE
1M 40Vpk
Signal Input Specifications
Item
Specifications
Connector type
BNC
Number of channels
2 (1 RF input connector and 1 clock input connector)
Input impedance
1 M
Ω
, 35 pF (typical value*)
Maximum input voltage
DC
≤
frequency of the input signal
≤
100 kHz:
40 V (DC+ACpeak)
100 kHz
≤
frequency of the input signal
≤
100 MHz:
{3.5/f + 5} V (DC+ACpeak), where f is a frequency in MHz.
Ground
Connect to the case ground
*
The typical value is a representative or standard value. It is not a warranted value.
CAUTION
Do not apply a voltage that exceeds the maximum input voltage to the input
connector. This may cause damage to the input section.
Note
When connecting the probe for the first time, perform phase correction of the probe according
to the description given in section 3.6. Failure to do so will cause unstable gain across
different frequencies, thereby preventing correct measurement. Phase correction of the
probe must be performed when the probe changes or the TA120E changes.