![YOKOGAWA 701924 User Manual Download Page 25](http://html.mh-extra.com/html/yokogawa/701924/701924_user-manual_911102025.webp)
13
IM 701924-01E
WARNING
As the frequency of the input signal increases, the maximum input voltage
of the probe decreases.
French
AVERTISSEMENT
Lorsque la fréquence du signal d'entrée augmente, la tension d'entrée
maximale de la sonde diminue.
Input voltage range
35V
–35V
35V
–35V
Use it within this range.
Vin+
Positive signal input terminal
Vin-
Negative signal input terminal
General Specifications
Standard power supply voltage*
±5 V ±5 %
Standard operating
environment
Temperature range
23 ± 5°C
Humidity range
55 ± 10%RH (no condensation)
Operating environment
Temperature range
5 to 40°C
Humidity range
20 to 80%RH (no condensation)
Altitude
2000 m or less
Storage environment
Temperature range
–20 to 60°C
Humidity range
20 to 80%RH (no condensation)
Altitude
3000 m or less
Warm-up time
30 minutes or more
Calibration period
1 year recommended
Total length
Approx. 1.2 m
Weight
Approx. 90 g
* Supplied through the interface pins from an oscilloscope with dedicated probe
interface.
Product Specifications